MedComp Recalls Tesio Catheter Sets Due to Connection Failure Risk
Medical Components, Inc. is recalling 712 Tesio catheter sets because connection failures could cause blood loss or air emboli.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (potential blood loss or air emboli) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Medical Components, Inc. (dba MedComp) is recalling 712 units of the 12F x 40cm Tesio Catheter Set with "Y" Adaptor (Reference CMD755, UDI 884908005530). The recall is due to a potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines.
The identified defect may contribute to blood loss or air emboli in patients. The affected products were distributed worldwide, including nationwide in the United States and in various international locations such as Australia, Canada, and several European countries.
Healthcare providers and consumers should stop using the affected catheter sets immediately. The specific lot numbers involved in this recall are MCBS820, MCBM670, MCDY280, MCDW670, MCFX 650, MCFW520, MCCC910, MCCJ210, MCDF970, MCCV540, MCFJ340, MCFQ090, and MCFS610.
The recalled product
- Product
- 12F X 40CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD755, UDI 884908005530 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
- Manufacturer
- Medical Components, Inc dba MedComp
- Affected units
- 712
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (12)
- Lot Numbers: MCBS820
- MCBM670
- MCDY280
- MCDW670
- MCFX 650
- MCFW520
- MCCC910
- MCCJ210
- MCDF970
- MCCV540
- MCFJ340
- MCFQ090 & MCFS610.
Distribution
Part of a larger recall action
This product is one of 10 recalled by Medical Components, Inc dba MedComp under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 10 products in this recall →Related recalls
Same manufacturer · Medical Components, Inc dba MedComp
See all →- ModerateMedComp Recalls Power Injectable CT Port Kits with Incorrect Port Size
FDA (Devices) · 2021-03-17
- ModerateMedcomp Titan HD Catheter Kits Recalled for Incorrect Valve Size
FDA (Devices) · 2020-09-16
- SevereMedComp C3 Wave App Recalled Due to ECG Display Failure
FDA (Devices) · 2019-11-06
- ModerateMedComp Recalls Hemodialysis Catheter Kits Due to Introducer Mismatch
FDA (Devices) · 2019-08-14
- ModerateMedComp Recalls Hemodialysis Catheter Kits Due to Labeling Error
FDA (Devices) · 2019-08-14
Same hazard · blood loss
See all →- SevereFresenius Medical Care Recalls 5008X Standard HD Pre-Flush Blood Tubing Sets
FDA (Devices) · 2026-09-09
- SevereFresenius 5008X HD/HDF Blood Tubing Set Recalled for Leak Risk
FDA (Devices) · 2026-09-09
- HighMidline Catheter Recalled Over Risk of Leaking or Cracking at Hub
FDA (Devices) · 2026-08-05
- HighPeripherally Inserted Central Catheters Recalled Due to Hub Leaking and Cracking Risk
FDA (Devices) · 2026-08-05
- SevereMedline Convenience Kits with Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08
Similar recalls
- HighMedComp Recalls Hemodialysis Catheter Kits Due to Suture Wing Cracking Risk
FDA (Devices) · 2018-09-05
- HighMedline apheresis and central line kits recalled over connector seal defects
FDA (Devices) · 2026-03-04
- HighMedComp Recalls Double Lumen CRRT Catheter Sets Due to Suture Wing Cracking
FDA (Devices) · 2018-09-05
- HighMedComp Recalls Duo-Flow Hemodialysis Catheter Sets Due to Suture Wing Cracking
FDA (Devices) · 2018-09-05
- HighMedComp Recalls Hemodialysis Catheter Kits Due to Suture Wing Cracking
FDA (Devices) · 2018-09-05
- ModerateMedComp Recalls Hemodialysis Catheter Kits Due to Labeling Error
FDA (Devices) · 2019-08-14