Philips HeartStart MRx Monitor/Defibrillator Recalled for ECG Cable Wear
Philips is recalling HeartStart MRx Monitor/Defibrillators because the ECG trunk cable and connector block may wear prematurely, causing an interrupted ECG signal.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device used to treat life-threatening cardiac arrhythmias. The hazard is an interrupted ECG signal, which is a risk-of-harm product issue where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Philips Medical Systems, Inc. is recalling approximately 49,929 units of the HeartStart MRx Monitor/Defibrillator. The affected models are M3535A, M3536A, M3536J, M3536M, M3536M5, and M3536MC. These devices are used for the termination of ventricular tachycardia and ventricular fibrillation.
The recall is due to a potential defect where the ECG trunk cable and connector block may be susceptible to accelerated wear. This wear could result in an interrupted ECG signal. The affected units have serial numbers within the range of US00100100 through US00550668.
The devices were distributed nationwide in the United States and in numerous countries worldwide. Healthcare facilities and users should check their device serial numbers against the affected range and contact Philips Medical Systems for further instructions on corrective actions.
The recalled product
- Product
- Philips HeartStart MRx Monitor/Defibrillator; Models: M3535A, M3536A, M3536J, M3536M, M3536M5, and M3536MC. Used for the termination of ventricular tachycardia and ventricular fibrillation.
- Manufacturer
- Philips Medical Systems, Inc.
- Affected units
- 49,929
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Serial numbers within the range: US00100100 through US00550668
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Philips Medical Systems, Inc.
See all →- ModeratePhilips Medical Systems Recalls Foot Switches for Bent Pedal Defect
FDA (Devices) · 2016-04-06
- ModeratePhilips MultiDiagnost Eleva Foot Switches Recalled for Bent Pedal Defect
FDA (Devices) · 2016-04-06
- SeverePhilips IntelliVue Information Center iX Software Recall for Alarm Failure
FDA (Devices) · 2015-12-16
- HighPhilips INTEGRIS Cardio System Ceiling Suspension Recall
FDA (Devices) · 2015-12-09
- HighPhilips Allura BiPlane Ceiling Suspension System Recall for Fall Risk
FDA (Devices) · 2015-12-09
Same hazard · signal interruption
See all →- HighBarco NexxisOR System Display Encoders and Decoders Recalled
FDA (Devices) · 2025-07-02