The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-1076-2021·Announced 2021-02-17

Argon Medical Devices Recalls Transjugular Liver Biopsy Systems Due to Sheath Separation

Argon Medical Devices is recalling 1,759 Transjugular Liver Biopsy Systems because a manufacturing error may cause the sheath to separate during use, potentially requiring surgical intervention.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The source text explicitly states that the defect could result in the need for surgical intervention to prevent permanent damage or impairment, which aligns with the rubric criteria for 'significant injury' or 'structural vehicle defect' (analogous to structural medical device defects) under the Severe category.

Plain-English summary

Argon Medical Devices, Inc. is recalling 1,759 units of the TLAB Transjugular Liver Biopsy System (REF: TL-18C). The affected lots are 1394204, 1424962, 1464709, 1473181, 1477229, 1481735, and 1495182. The product was distributed to various states in the United States and to several countries internationally, including Canada, Germany, and Italy.

The recall is due to a manufacturing error that creates a potential for the sheath within the biopsy kit to separate during use if the tip is flexed or experiences bending stresses. This defect could result in the need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.

Healthcare providers and facilities that have received these devices should stop using them and contact Argon Medical Devices for further instructions on the recall process.

The recalled product

Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 18C, 1/18 GA Flexcore Biopsy Needle, 1/7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, 1/5f straight catheter, 1/5F Curved Catheter Sterile EO,
Affected units
1,759

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (7)

  • Lot Number: 1394204
  • 1424962
  • 1464709
  • 1473181
  • 1477229
  • 1481735
  • 1495182

Distribution

Part of a larger recall action

This product is one of 6 recalled by Argon Medical Devices, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 6 products in this recall →

Similar recalls