Daavlin 1 Series CX Phototherapy Units Recalled for Software Dosing Error
Daavlin is recalling 1 Series CX phototherapy units due to a software bug that swaps calibration distance settings when resuming interrupted treatments, resulting in incorrect dosing.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a risk-of-harm medical device where a confirmed software defect creates potential for incorrect dosing. No illnesses or injuries have been reported in the source, and FDA Class II recalls without reported harm typically score at 3 per the rubric.
Plain-English summary
Daavlin Distributing Company is recalling 1 Series CX phototherapy units equipped with ClearLink Control System software versions v3.03h and 3.04v. A total of 216 units were distributed worldwide, including across most US states, Australia, and Canada.
The recall addresses a software defect that occurs when resuming an interrupted treatment. The software swaps the calibration treatment distance from 0 inches to 9 inches or vice versa, resulting in over or under dosing.
The affected models are 1 Series CX 305-4, 1 Series CX 311-4, 1 Series CX 350-4, and 1 Series CX UVA1-4 (all 120-240V 50/60Hz configurations). Healthcare facilities and practitioners should verify their units against the provided serial numbers and UDI codes in the FDA recall notice.
The recalled product
- Product
- 1 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v. Models: [1 Ser CX 305-4 120-240V 50/60Hz, item # 06OS0004CX6], [1 Ser CX 311-4 120-240V 50/60Hz CE, item # 807OS0004CX5], [1 Ser CX 311-4 120-240V 50/60Hz, item…
- Manufacturer
- Daavlin Distributing Company
- Affected units
- 216
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI-DI: +B2809806OS0004CX60K
- +B2809807OS0004CX50K
- +B2809807OS0004CX60L
- +B2809807OS0400CX60L
- and +B2809807OS0400CX6A10W. Serial Numbers: OS22183PLNBCX0004
- OS22320PLNBCX0004
- OS23725PLNBCX0004
- OS24083PLNBCX0004
- OS24089PLNBCX0004
- OS24090PLNBCX0004
- OS23532PLA1CX0400
- OS22097PLNBCX0004
- OS22200PLNBCX0004
- OS22265PLNBCX0004
- OS22270PLNBCX0004
- OS22304PLNBCX0004
- OS22739PLNBCX0004
- OS22788PLNBCX0004
- OS22795PLNBCX0004
- OS23239PLNBCX0004
Distribution
Part of a larger recall action
This product is one of 2 recalled by Daavlin Distributing Company under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Daavlin Distributing Company
See all →- ModeratePhotera 600 3-D phototherapy device UV sensor calibration error recall
FDA (Devices) · 2026-07-22
- ModeratePhotera 400 Phototherapy Device Recalled for Incorrect UV Sensor Calibration
FDA (Devices) · 2026-07-22
- ModeratePhotera 600 Phototherapy Device UV Sensor Calibration Error
FDA (Devices) · 2026-07-22
- HighPhothera 200 phototherapy calibration certificates had outputs reversed
FDA (Devices) · 2026-06-10
- HighPhototherapy Units Recalled for Software Bug Causing Incorrect Dosing
FDA (Devices) · 2022-05-18
Same hazard · software defect
See all →- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uMI 550 PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModeratePhilips SmartPath to MR 7700 Recalled for Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
Similar recalls
- ModeratePhotera 600 3-D phototherapy device UV sensor calibration error recall
FDA (Devices) · 2026-07-22
- ModeratePhotera 400 Phototherapy Device Recalled for Incorrect UV Sensor Calibration
FDA (Devices) · 2026-07-22
- ModeratePhotera 600 Phototherapy Device UV Sensor Calibration Error
FDA (Devices) · 2026-07-22
- SevereBrainlab ExacTrac Dynamic Radiation Therapy System Software Recall
FDA (Devices) · 2021-03-17
- HighMobile X-Ray System Recalled for Software Calibration Error in Radiation Dose Meter
FDA (Devices) · 2022-10-05
- HighRayStation Radiation Therapy Software Recalled for Potential Dose Errors
FDA (Devices) · 2014-04-09