Integra Ojemann Cortical Stimulator Recalled for Potential Electrical Injury Risk
Integra Burlington MA, Inc. is recalling 371 Ojemann Cortical Stimulators because a non-intended voltage could be transmitted to patients during surgery.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a potential risk of injury to patients, but the source text does not report any actual injuries or illnesses, fitting the criteria for a high-risk product where injury has not yet been reported.
Plain-English summary
Integra Burlington MA, Inc. is recalling 371 units of the Integra Ojemann Cortical Stimulator (Catalogue No. OCS2). The device is intended for intraoperative cortical stimulation mapping procedures. The affected units have serial numbers ranging from 1010 through 1380.
The recall is due to a potential safety hazard where a non-intended voltage, such as static electricity, could be transmitted to the patient if the headphone jack is in use during a surgical procedure and comes in contact with the outer case of the unit. This transmission could be a potential source of injury.
The affected devices were distributed worldwide, including nationwide in the United States and in Austria, Belgium, Czech Republic, Denmark, Spain, Great Britain, Hungary, Ireland, Italy, Netherlands, Poland, Morocco, Portugal, Saudia Arabia, Sweden, Turkey, and South Africa. Users should contact the manufacturer for instructions on how to proceed.
The recalled product
- Product
- Integra Ojemann Cortical Stimulator; Product / Catalogue No: OCS2. Intended for intraoperative cortical stimulation mapping procedures.
- Manufacturer
- Integra Burlington MA, Inc.
- Hazard
- Affected units
- 371
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Serial Number Range: 1010 through 1380
Distribution
Distributed nationwide across the United States.
Related recalls
Same hazard · electrical injury
See all →- HighAlpha Omega Engineering Recalls Neuro Omega Systems Due to Tissue Harm Risk
FDA (Devices) · 2019-05-15
Other medical device recalls
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighMedline Recalls 71 Convenience Kits Due to Contaminated Component
FDA (Devices) · 2026-09-02
- HighMedline Recalls 419 Surgical Kits Due to Contaminated Component
FDA (Devices) · 2026-09-02
- ModerateMedline Trach Tube Kit Recalled for Component Labeling Discrepancy
FDA (Devices) · 2026-09-02