Brainlab VectorVision Spine Instrument Star Unit ML Recall
Brainlab is recalling 25 VectorVision Spine Instrument Star Unit ML devices because pins may be made of incorrect material, posing a risk of corrosion and particle release during surgery.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where a material defect could lead to particle release during surgery, representing a risk of harm where injury has not yet been reported.
Plain-English summary
Brainlab AG is recalling 25 units of the VectorVision Spine Instrument Star Unit ML (Model/Catalog Code: 55830-25A). The recall is due to a determination that pins in the Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material. This material is not suitable for the necessary disinfection and sterilization of the product.
The affected devices are intended for use as an intraoperative image-guided localization system for minimally invasive surgery. The system links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data. The recall covers devices distributed worldwide, including in the United States (CA, MN, NC, NY, OH, TX, and DC) and internationally to Australia, Chile, Germany, Hungary, Italy, Panama, Qatar, Russia, Thailand, and the United Arab Emirates.
If corrosion occurs on the pins and is not detected by the user, and the device is used during surgery, corroded particles of the pin could be released directly or indirectly. Users should stop using the affected devices and contact Brainlab AG for further instructions.
The recalled product
- Product
- BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data…
- Manufacturer
- Brainlab AG
- Affected units
- 25
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model/Catalog Code: 55830-25A Instrument Star Unit ML
- Serial Numbers (US Devices): 1205112007
- 1205112010
- 1205112011
- 1205112013
- 1205112014
- 1205112019
- 1205112021
- 1205112029
- 1205112034
- 1205112045
- Serial Numbers (Non-US Devices): 1205112004
- 1205112006
- 1205112008
- 1205112009
- 1205112015
- 1205112016
- 1205112017
- 1205112018
- 1205112020
Distribution
Related recalls
Same manufacturer · Brainlab AG
See all →- HighMedical Device Software May Merge Patient Records Without Notification
FDA (Devices) · 2025-03-26
- HighSurgical navigation system recalled for potential mechanical instability
FDA (Devices) · 2023-04-05
- HighRadiation Therapy Software May Not Properly Execute Heart-Protection Breath Hold
FDA (Devices) · 2023-03-22
- HighExacTrac Dynamic Software Yaw Angle Error May Affect Radiotherapy Positioning
FDA (Devices) · 2022-03-16
- HighRadiotherapy positioning software may delay patient movement display
FDA (Devices) · 2021-06-16
Same hazard · corrosion
See all →- HighToyota 2024 Tacoma rear shock absorber suspension recall issued
NHTSA · 2026-08-06
- High
- HighArvis Hip and Knee 2.0 trackers may show wear and corrosion
FDA (Devices) · 2026-06-10
- HighPhaseOne antimicrobial solution sprayer springs may corrode into the product
FDA (Devices) · 2026-06-10
- HighArvis Hip and Knee 3.0 trackers may show wear and corrosion
FDA (Devices) · 2026-06-10
Similar recalls
- HighBrainlab Navigation Spine & Trauma 3D Software Recalled for Registration Error
FDA (Devices) · 2014-08-20
- SevereBrainlab Spine & Trauma 3D Navigation Software Recall for Display Error
FDA (Devices) · 2019-04-24
- HighBrainlab Spine & Trauma 3D 2.0 Navigation Software Recalled for Registration Defect
FDA (Devices) · 2014-04-23
- SevereBrainlab Cranial Navigation System Recalled for Potential Software Inaccuracy
FDA (Devices) · 2016-01-27
- ModerateBrainlab Ultrasound Navigation Software Recalled for Image Width Limitation
FDA (Devices) · 2020-10-14
- HighBrainlab Offset Cup Impactor Universal Recall Due to Sterilization Failure
FDA (Devices) · 2015-03-04