Xeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
Xeridiem is recalling 37 EndoVive 3s Low Profile Balloon Kits because the feeding valve may close inconsistently, allowing fluid to leak.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device defect that allows leakage of gastric contents or feeding solution, representing a risk of harm where injury has not yet been reported.
Plain-English summary
Xeridiem Mediem Medical Devices Inc is recalling 37 units of the EndoVive 3s Low Profile Balloon Kits, Part Number M00548870 (XMD P/N 70-0050-620), Lot Number 929729. The recall is limited to nationwide distribution to Massachusetts only.
The recall is due to a feeding (reflux) valve leakage issue. The feeding valve may close inconsistently or in a delayed manner when an extension set is disconnected from the device. This defect allows feeding solution or gastric contents to leak out of the device in some instances.
The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration, and gastric decompression. It is indicated for adult and pediatric patients who require long-term feeding, are unable to tolerate oral feeding, are at low risk for aspiration, or require gastric decompression and/or medication delivery directly into the stomach.
The recalled product
- Product
- EndoVive 3s Low Profile Balloon Kits Part Number: M00548870 (XMD P/N 70-0050-620) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The…
- Manufacturer
- Xeridiem Mediem Medical Devices Inc
- Affected units
- 37
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Part Number: M00548870 (XMD P/N 70-0050-620) Lot Number: 929729
Distribution
Part of a larger recall action
This product is one of 105 recalled by Xeridiem Mediem Medical Devices Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 105 products in this recall →Related recalls
Same manufacturer · Xeridiem Mediem Medical Devices Inc
See all →- HighXeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Balloon Kits Recalled for Feeding Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
Same hazard · leakage
See all →- SevereDraeger VentStar Resus Neo Neonatal Breathing Circuit Recall
FDA (Devices) · 2026-10-07
- HighConvaTec Esteem Body Soft Convex Urostomy Pouches Recalled for Leakage
FDA (Devices) · 2026-08-26
- SevereStryker AHTO Tube Set Packaging Recalled for Handpiece Leaking
FDA (Devices) · 2026-08-12
- SevereMedtronic Perfusion Systems Recalls EOPA Elongated One-Piece Arterial Cannula Model 77522
FDA (Devices) · 2026-08-05
- HighICU Medical Recalls Hotline Fluid Warming Sets Due to Leakage Risk
FDA (Devices) · 2026-07-29
Similar recalls
- HighEndoVive 20 Fr Feeding Tube May Obstruct During Placement
FDA (Devices) · 2024-04-17
- HighEndoVive 20Fr Push Safety PEG Kit Recall Due to Blocked Lumen Risk
FDA (Devices) · 2024-04-17
- SevereEZDilate Wire Guided Balloon Recalled for Device Malfunction and Patient Injury Risk
FDA (Devices) · 2024-02-07
- SevereEZDilate Wire Guided Balloon Recalled for Device Malfunction Issues
FDA (Devices) · 2024-02-07
- SevereEndoscopic Dilation Balloons Recalled for Bursting and Leaking Failures
FDA (Devices) · 2024-02-07
- HighPEG Feeding Tube Kit Recalled for Blocked Lumen Preventing Guidewire Advancement
FDA (Devices) · 2024-04-17