The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0943-2016·Announced 2016-03-02

Xeridiem EndoVive 3s Balloon Kits Recalled for Valve Leakage

Xeridiem Mediem Medical Devices is recalling 80 EndoVive 3s Low Profile Balloon Kits because the feeding valve may close inconsistently, allowing fluid to leak.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device defect that poses a risk of harm (leakage of gastric contents or feeding solution) without reported injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Xeridiem Mediem Medical Devices Inc is recalling 80 units of the EndoVive 3s Low Profile Balloon Kits, Part Number M00548700 (XMD P/N 70-0050-518). The affected lots are 928921 and 929711. These devices are intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration, and gastric decompression in adult and pediatric patients.

The recall is due to a feeding (reflux) valve leakage issue. The feeding valve may close inconsistently or in a delayed manner when an extension set is disconnected from the Low Profile Balloon device. This malfunction can allow feeding solution or gastric contents to leak out of the device in some instances.

The products were distributed nationwide to Massachusetts only. Patients using these devices should contact their healthcare provider or the manufacturer for instructions on how to proceed.

The recalled product

Product
EndoVive 3s Low Profile Balloon Kits Part Number: M00548700 (XMD P/N 70-0050-518) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The…
Affected units
80

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Part Number: M00548700 (XMD P/N 70-0050-518) Lot Number: 928921 929711

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Part of a larger recall action

This product is one of 105 recalled by Xeridiem Mediem Medical Devices Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 105 products in this recall →

Similar recalls