Xeridiem EndoVive 3s Balloon Kits Recalled for Valve Leakage
Xeridiem is recalling 323 EndoVive 3s Low Profile Balloon Kits because the feeding valve may leak gastric contents or feeding solution.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a functional defect that poses a risk of harm (leakage of gastric contents/feeding solution), but the source text does not report any specific injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Xeridiem Mediem Medical Devices Inc is recalling 323 units of the EndoVive 3s Low Profile Balloon Kits, Part Number M00548370 (XMD P/N 70-0050-414). The affected lots are 911108, 915408, 915908, 920608, 928909, and 930839. The devices were distributed nationwide to Massachusetts only.
The recall is due to a feeding (reflux) valve leakage issue. The valve may close inconsistently or with a delay when an extension set is disconnected from the device. This can allow feeding solution or gastric contents to leak out of the device.
The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration, and gastric decompression in adult and pediatric patients. Patients using these devices should contact their healthcare provider for instructions on how to proceed.
The recalled product
- Product
- EndoVive 3s Low Profile Balloon Kits Part Number: M00548370 (XMD P/N 70-0050-414) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The End
- Manufacturer
- Xeridiem Mediem Medical Devices Inc
- Affected units
- 323
Distribution
Part of a larger recall action
This product is one of 105 recalled by Xeridiem Mediem Medical Devices Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 105 products in this recall →Related recalls
Same manufacturer · Xeridiem Mediem Medical Devices Inc
See all →- HighXeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Balloon Kits Recalled for Feeding Valve Leakage
FDA (Devices) · 2016-03-02
- HighXeridiem EndoVive 3s Low Profile Balloon Kits Recalled for Valve Leakage
FDA (Devices) · 2016-03-02
Same hazard · leakage
See all →- HighConvaTec Esteem Body Soft Convex Urostomy Pouches Recalled for Leakage
FDA (Devices) · 2026-08-26
- SevereStryker AHTO Tube Set Packaging Recalled for Handpiece Leaking
FDA (Devices) · 2026-08-12
- SevereMedtronic Perfusion Systems Recalls EOPA Elongated One-Piece Arterial Cannula Model 77522
FDA (Devices) · 2026-08-05
- HighICU Medical Recalls Hotline Fluid Warming Sets Due to Leakage Risk
FDA (Devices) · 2026-07-29
- SevereABIOMED Introducer Kit 23 Fr heart pump accessory leakage risk
FDA (Devices) · 2026-07-22