The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0931-2016·Announced 2016-03-02

Xeridiem EndoVive 3s Balloon Kits Recalled for Feeding Valve Leakage

Xeridiem is recalling 32 EndoVive 3s Low Profile Balloon Kits because the feeding valve may close inconsistently, allowing fluid to leak.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a medical device defect that allows leakage of gastric contents or feeding solution, which poses a risk of harm to patients, but no specific injuries or illnesses are reported in the source text.

Plain-English summary

Xeridiem Mediem Medical Devices Inc is recalling 32 units of the EndoVive 3s Low Profile Balloon Kits (Part Number M00549330, Lot Number 930140). The recall is due to a leakage issue with the feeding (reflux) valve.

The feeding valve may close inconsistently or in a delayed manner when an extension set is disconnected from the device. This defect allows feeding solution or gastric contents to leak out of the device in some instances. The product is intended to provide gastric access for enteral feeding, medication administration, and gastric decompression in adult and pediatric patients.

The affected units were distributed nationwide to Massachusetts only. Patients and caregivers should contact their healthcare provider or the manufacturer for instructions on how to proceed with the recall.

The recalled product

Product
EndoVive 3s Low Profile Balloon Kits Part Number: M00549330 (XMD P/N 70-0050-322) The EndoVive 3s Low Profile Balloon is intended to provide gastric access through a gastrointestinal stoma tract for enteral feeding, medication administration and gastric decompression. The…
Affected units
32

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Part Number: M00549330 (XMD P/N 70-0050-322) Lot Number: 930140

Distribution

Distributed nationwide across the United States.

The notice also names 1 state specifically:

Part of a larger recall action

This product is one of 105 recalled by Xeridiem Mediem Medical Devices Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 105 products in this recall →