Medtronic Harmony Delivery Catheter System Recalled Over Equipment Failure Risk
Medtronic is recalling 1,483 Harmony Delivery Catheter Systems due to potential capsule bond failure during pulmonary valve replacement procedures. The FDA has classified this as a Class I recall affecting devices distributed in the US and Canada.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: This device recall has been classified as FDA Class I, the most serious FDA category. Under the severity rubric, FDA Class I recalls require a minimum severity score of 4. While no deaths or reported injuries are documented in the source, the Class I designation reflects the serious nature of a potential equipment failure during a cardiac procedure.
Plain-English summary
Medtronic Inc is recalling the Harmony Delivery Catheter System, which is part of the Harmony Transcatheter Pulmonary Valve (TPV) Replacement Implantation System. The recall affects 1,483 devices distributed nationwide in the United States and in Canada.
The FDA has classified this as a Class I recall. The hazard involves a potential defect in which the capsule bond may break during the procedure. If this occurs, it could affect the proper delivery and placement of the valve replacement implant.
Healthcare providers who have received any of the affected lot numbers should immediately stop using the devices and contact Medtronic. Patients who have received this device, or who are scheduled for a procedure using this system, should speak with their healthcare provider about the recall.
The recalled product
- Product
- Harmony Delivery Catheter System. Part of the Harmony Transcatheter Pulmonary Valve (TPV) Replacement Implantation System.
- Manufacturer
- Medtronic Inc
- Affected units
- 1,483
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- HARMONY-DCS (GTIN: 00763000341367)
- Lot Numbers: 0010519040
- 0010565921
- 0010578123
- 0010578126
- 0010583884
- 0010631963
- 0010631969
- 0010631971
- 0010717000
- 0010717001
- 0010717002
- 0010723188
- 0010723189
- 0010723190
- 0010727345
- 0010804938
- 0010826589
- (the following lots were added 4/22/2022) 0010826591
- 0010826592
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Medtronic Inc
See all →- HighMedtronic Sphere-9 Catheter may cause ventricular fibrillation with Biotronik ICDs
FDA (Devices) · 2026-05-06
- HighVenous stents recalled over a nitinol anomaly that reduces deployed diameter
FDA (Devices) · 2026-01-21
- ModerateKidney tests recalled for distribution without premarket approval or clearance
FDA (Devices) · 2025-12-10
- HighSmartSync Patient Connector limited the window to abort high voltage therapy
FDA (Devices) · 2025-07-30
- HighCareLink SmartSync limited the window to abort a high voltage therapy
FDA (Devices) · 2025-07-30
Same hazard · equipment failure
See all →- High
- HighHeidelberg SPECTRALIS Flex Module Recall Due to Wheel Detachment Risk
FDA (Devices) · 2026-09-30
- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- SevereOmnipod Eros Pod Insulin Delivery Failure Risk
FDA (Devices) · 2026-07-08
Similar recalls
- SevereMedtronic Harmony Delivery Catheter System Distal Tip Detachment Risk
FDA (Devices) · 2026-07-08
- HighMedtronic recalls C315-HIS Delivery Catheters due to functional defect
FDA (Devices) · 2021-11-03
- SevereImplantable Cardiac Devices Risk Therapy Failure From Feedthrough Defect
FDA (Devices) · 2023-06-28
- CriticalFDA Recalls HeartWare HVAD Pump Implant Kit for Neurological Events and Mortality
FDA (Devices) · 2021-08-18
- SevereMedtronic Implantable Defibrillators Risk Failure to Deliver High-Voltage Therapy
FDA (Devices) · 2023-06-28
- SevereMedtronic Cardiac Defibrillators Potentially Fail to Deliver High-Voltage Therapy
FDA (Devices) · 2023-06-28