Agency classClass IIOur severity:SevereFDA (Devices)·Z-0922-2017·Announced 2017-02-01

Philips Allura Xper FD20 C-arm Recall for Loose Screws

Philips is recalling one Allura Xper FD20 unit because production errors left screws loose, which could cause the C-arm to rotate uncontrolled.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall involves a structural defect in a medical device where uncontrolled rotation of the C-arm could cause significant injury, meeting the criteria for a Severe rating.

Plain-English summary

Philips Electronics North America Corporation is recalling one unit of the Allura Xper FD20 722012 medical device. This device is intended for cardiovascular and vascular X-ray imaging applications, including diagnostic and interventional procedures such as PTCA, stent placement, atherectomies, pacemaker implantations, and Electro Physiology (EP).

The recall is due to a production error where screws were not tightened to the specified torque. These screws may come loose from the C-arm shaft, causing the C-arm to rotate uncontrolled.

The affected unit was distributed in Washington, China, Spain, Korea, Republic of, and Germany. Users of this device should contact Philips Electronics North America Corporation for further instructions on how to address the issue.

The recalled product

Product
Allura Xper FD20 722012 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).
Affected units
1

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • 2786

Distribution

Distributed in 1 state:

Part of a larger recall action

This product is one of 4 recalled by Philips Electronics North America Corporation under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 4 products in this recall →

Same manufacturer · Philips Electronics North America Corporation

See all →

Same hazard · uncontrolled rotation

See all →