The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0920-2013·Announced 2013-03-20

Integra CUSA CEM Nosecone Recalled for Potential Tip Erosion

Integra LifeSciences is recalling CEM Nosecones for the CUSA EXcel system due to a potential for tip erosion and breakage. No patient injuries have been reported.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a risk of harm (tip breakage) where injury has not yet been reported, which aligns with the 'High' severity criteria in the rubric.

Plain-English summary

Integra LifeSciences Corp. is recalling CEM Nosecones for the CUSA EXcel System. The recall involves specific lot numbers of Catalogue No. C6623 and C6636. The affected units were distributed worldwide, including in the United States, Europe, and other international markets.

The recall is issued because there is a potential for erosion of the CUSA EXcel tip when used with the Nosecone. This erosion can occur if the side of the tip, rather than the annulus, is in contact with tissue during use. The erosion can potentially lead to tip breakage.

There were no reports of patient injury associated with these complaints. The source text does not provide specific instructions for consumers or healthcare providers on how to handle the recalled devices.

The recalled product

Product
Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined…

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • CEM Nosecone Catalogue No. C6623 Lot numbers: 1092855
  • 1093129
  • 1100436
  • 1100437
  • 1101217
  • 1101328
  • 1103393
  • 1104265
  • 1105067
  • 1110311
  • 1110892
  • 1112182
  • 1112183
  • 1114112
  • 1114655
  • 1115225
  • 1115499
  • 1115500
  • 1120771
  • 1121141

Distribution

Distribution scope not specified by the agency.