Radiation therapy planning software: incorrect dose density calculation
RayStation/RayPlan radiation therapy software (versions 4–11A) may incorrectly calculate radiation dose density in treatment planning. The error affects dose calculations in regions partially covered by certain structures.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The calculation error affects critical dose determinations in radiation therapy planning, but the source text reports no actual injuries, hospitalizations, or adverse events. Per the FDA severity rubric, a risk-of-harm product without reported injury qualifies as High. Class II classification applies.
Plain-English summary
The FDA has recalled RayStation/RayPlan versions 4 through 11A, software used in healthcare facilities to calculate and plan radiation therapy treatments. The recall affects 478 licensed copies distributed nationwide in the United States.
The software contains an error in dose grid voxel density calculations. When voxels are partially covered by certain regions of interest (ROIs)—specifically Bolus, Support, or Fixation regions—the density values may be incorrectly estimated. Affected voxels can show either underestimated or overestimated density, potentially leading to inaccurate radiation dose calculations in treatment plans.
Healthcare facilities using RayStation/RayPlan versions listed in the recall should contact RaySearch Laboratories AB for corrected software versions. Treatment plans created with affected software should be reviewed to verify dose calculation accuracy. Consult the FDA recall notice for specific build version numbers and UDI codes for each affected product version.
The recalled product
- Product
- RayStation/RayPlan versions 4 - 11A, including service packs. Build Stations: 4.0.0.14, 4.0.3.4, 4.3.0.14, 4.5.1.14, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.7.6.7, 4.9.0.42, 5.0.1.11, 5.0.2.35, 5.0.3.17, 6.0.0.24, 6.1.1.2, 6.2.0.7, 6.3.0.6, 7.0.0.19, 8.0.0.61, 8.0.1.10…
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Affected units
- 478
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · dose calculation error
See all →- HighRadiation Therapy Planning System Density Calculation Inconsistency in RayStation
FDA (Devices) · 2025-02-26
- HighRayStation radiation therapy planning software density uncertainty calculation inconsistency
FDA (Devices) · 2025-02-26
- HighMedical device recall: RayStation radiation therapy software dose calculation inconsistency
FDA (Devices) · 2025-02-26
- HighRayStation Radiation Therapy Planning Software Recall for Density Calculation Inconsistency
FDA (Devices) · 2025-02-26
- HighRadiation Therapy Planning Software Correction for Density Uncertainty in Dose Calculations
FDA (Devices) · 2025-02-26
Similar recalls
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Errors
FDA (Devices) · 2016-12-14
- HighRadiation Therapy Planning System Density Calculation Inconsistency in RayStation
FDA (Devices) · 2025-02-26
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Accuracy Issues
FDA (Devices) · 2018-07-04
- HighMedical software recall: RayStation radiation therapy dose calculation errors
FDA (Devices) · 2022-04-06
- HighRayStation radiation therapy software recall for density calculation inconsistency
FDA (Devices) · 2025-02-26
- HighRadiation therapy planning system doses may be calculated incorrectly
FDA (Devices) · 2024-05-08