The Recall Desk
HighFDA (Devices)·Z-0797-2016·Announced 2016-02-17

Orthofix Firebird Spinal Fixation System Torque Limiting Handle Recall

Orthofix is recalling 36 Firebird Spinal Fixation System Torque Limiting Handles because they may deliver less torque than required after sterilization, potentially causing spinal construct loosening.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (spinal construct loosening requiring corrective surgery) where injury has not yet been reported, fitting the 'High' severity criteria.

Plain-English summary

Orthofix, Inc. is recalling 36 units of the Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512). The device is intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium). The recall covers units distributed worldwide, including the US, Spain, Italy, and Australia.

The recall is due to a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In the worst case, this unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.

The source text does not specify specific actions for consumers or healthcare providers, nor does it report any specific illnesses or injuries associated with this recall.

The recalled product

Product
Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512) Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).
Manufacturer
Orthofix, Inc
Affected units
36

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • BS535418105
  • BS535419118
  • BS535420030
  • BS535419124
  • BS562325021
  • BS535418158
  • BS535418038
  • BS535418138
  • BS535418012
  • BS535419006
  • BS535420039
  • BS535418127
  • BS549024010
  • BS535419141
  • BS535418081
  • BS535418111
  • BS535418117
  • BS535420041
  • BS535418144
  • BS535420081

Distribution

Distribution scope not specified by the agency.