Beckman Coulter Recalls MicroScan LabPro Data Management System Software
Beckman Coulter is recalling MicroScan LabPro Data Management System version 4.42 due to intermittent behavior that may delay reporting results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text describes a workflow interruption and potential delay in reporting results without mentioning any reported injuries, illnesses, or hospitalizations. This aligns with the 'Moderate' criteria for low-risk issues or voluntary precautionary recalls where harm has not been reported.
Plain-English summary
Beckman Coulter Inc. is recalling the MicroScan LabPro Data Management System, software intended to manage microbial identification and antimicrobial agent susceptibility testing data for use by trained laboratory personnel. The recall affects 131 units total, with 130 units distributed in the United States and the remainder in Mexico. The specific software version involved is 4.42, associated with part numbers 6000-0050 through 6000-0054 and 6000-0060.
The recall was initiated after Beckman Coulter received and confirmed reports of intermittent and unexpected behavior when loading new panels into a WalkAway instrument. This issue could cause workflow interruptions and potentially delay the reporting of results because the system may be unable to begin processing new panels.
The source text does not specify the corrective action or instructions for affected users. Affected laboratory personnel should contact Beckman Coulter for further information regarding the recall and any necessary remediation steps.
The recalled product
- Product
- MicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory…
- Manufacturer
- Beckman Coulter Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (7)
- Software Version 4.42
- Part No. 6000-0050
- 6000-0051
- 6000-0052
- 6000-0053
- 6000-0054
- and 6000-0060.
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Beckman Coulter Inc.
See all →- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- HighBeckman Coulter DxC 500i Analyzer Modules Recalled for Software Defect
FDA (Devices) · 2026-08-26
- HighBeckman Coulter DxC 500 AU Analyzer Software Defect Recall
FDA (Devices) · 2026-08-26
- High
Same hazard · software defect
See all →- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uCT 550 CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uMI Panvivo S PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModeratePhilips SmartPath to MR 7700 Recalled for Incorrect DTI Export Data
FDA (Devices) · 2026-10-07
Similar recalls
- ModerateBeckman Coulter LabPro Data Management System Software Update Recall
FDA (Devices) · 2020-04-15
- HighBeckman Coulter MicroScan LabPro Software Recall for Manual Editing Error
FDA (Devices) · 2015-09-30
- ModerateBeckman Coulter Recalls MicroScan LabPro Software Due to Editing Error
FDA (Devices) · 2016-04-06
- SevereBeckman Coulter MicroScan Panel Update DVD Kits Recalled
FDA (Devices) · 2026-08-12
- High
- ModerateBeckman Coulter Recalls MicroScan Panels Due to Performance Limitation
FDA (Devices) · 2015-05-06