Merlin PCS 3650 Pacemaker Programmer Software Anomaly May Cause Loss of Pacing
A software anomaly in Abbott's Merlin PCS 3650 programmer may prevent proper finalization of leadless pacemakers, causing loss of pacing function. The issue occurs when specific actions combine with brief telemetry loss.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II device recall involving a potential loss-of-function defect in a critical life-sustaining medical device. While no illnesses or injuries have been reported in the source text, loss of pacing function in a pacemaker could cause serious harm. Per the rubric, this constitutes a 'risk-of-harm product where injury has not yet been reported,' warranting a High (3) rating.
Plain-English summary
The Merlin PCS 3650 programmer is a portable device used to program and monitor implantable leadless pacemakers manufactured by Abbott Medical. Due to a software anomaly, a specific sequence of programmer actions combined with a brief loss of telemetry (less than 2 seconds) may cause the finalization step of the pacemaker setup to fail.
When this occurs, the programmer displays a 'loss of telemetry detected' message and the pacemaker stops pacing. This condition could result in inadequate cardiac output and serious patient harm.
Approximately 7,295 programmers have been distributed worldwide, including throughout the United States and multiple international locations. Affected patients should contact their healthcare provider or Abbott Medical to determine if their device is affected and discuss monitoring or management options.
The recalled product
- Product
- The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical…
- Manufacturer
- Abbott Medical
- Affected units
- 7,295
Distribution
Related recalls
Same manufacturer · Abbott Medical
See all →- HighAbbott Merlin pacemaker programmer may fail to terminate threshold test
FDA (Devices) · 2026-06-17
- HighDeep brain stimulators recalled over loss of programmer communication
FDA (Devices) · 2025-11-19
- HighSpinal cord stimulators recalled over loss of programmer communication
FDA (Devices) · 2025-11-19
- HighAbbott Aveir DR Leadless Pacemaker Recalled for Battery Failure Risk
FDA (Devices) · 2025-04-23
- ModerateHeart Device Transmitter Fails to Update Software After Power Interruption
FDA (Devices) · 2024-12-04
Same hazard · software failure
See all →- ModeratePhilips MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModeratePhilips MRI Software Recall for Image Reconstruction Failure
FDA (Devices) · 2026-09-30
- ModerateFoundation Medicine Recalls Hamilton Microlab STAR Liquid Handler Software
FDA (Devices) · 2026-08-26
- SevereOlympus High Flow Insufflation Unit Recalled Due to Software Overpressure Risk
FDA (Devices) · 2026-02-25
- ModerateStryker Blueprint Mixed Reality Glenoid Pin Guide Impaired Function Recall
FDA (Devices) · 2025-04-16
Similar recalls
- HighMerlin PCS 3650 Programmer May Fail to Terminate Pacing Test Properly
FDA (Devices) · 2022-04-27
- HighAbbott Merlin pacemaker programmer may fail to terminate threshold test
FDA (Devices) · 2026-06-17
- HighPacemaker Programmer Software May Display Overestimated Battery Longevity
FDA (Devices) · 2022-08-10
- SevereAbbott Recalls Merlin PCS 3650 Software Upgrade Kit Due to Anomaly
FDA (Devices) · 2020-04-22
- HighSt. Jude Medical Recalls Merlin PCS 3650 Programmer for Firmware Update
FDA (Devices) · 2017-10-25
- HighCardiac Device Programmer Software May Fail to Terminate Test
FDA (Devices) · 2022-04-27