HeartWare Model 1102 Driveline Cover May Harden, Delaying Critical Connector Access
HeartWare Model 1102 cardiac assist devices are being recalled because the driveline cover may harden over time, preventing timely access to the critical driveline connector during urgent clinical situations.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Class II medical device with a defect preventing access to critical cardiac connectors during emergencies. Although no hospitalizations have been reported, the potential for serious patient harm from prolonged pump cessation in cardiac patients meets the criterion for high-severity risk-of-harm products.
Plain-English summary
The HeartWare Model 1102 cardiac assist device is being recalled due to potential hardening of the driveline cover over time. This hardening may make the cover difficult or impossible to slide back, preventing access to the driveline-to-controller connector.
Access to the driveline connector is critical for addressing urgent device issues. Delayed access caused by a hardened driveline cover may result in extended pump stoppage and potential patient harm during emergency situations.
The recall affects 892 units distributed worldwide. Patients and healthcare providers using affected devices should contact a Medtronic field representative for assessment and servicing if they observe a hardened driveline cover.
The recalled product
- Product
- HeartWare Model 1102
- Manufacturer
- Heartware, Inc.
- Affected units
- 892
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Serial Numbers: HW951
- HW1000
- HW1001
- HW1002
- HW1003
- HW1006
- HW1007
- HW1008
- HW1010
- HW1011
- HW1013
- HW1017
- HW1018
- HW1019
- HW1026
- HW1027
- HW1029
- HW1030
- HW1034
- HW1035
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 9 recalled by Heartware, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 9 products in this recall →Related recalls
Same manufacturer · Heartware, Inc.
See all →- HighHVAD Ventricular Assist Device Instructions Updated for Controller Alarm and Lifespan
FDA (Devices) · 2023-12-13
- ModerateHeartWare HVAD Implant Kit Instructional Update for Fault Alarms
FDA (Devices) · 2023-12-13
- HighHVAD Ventricular Assist Pump Model 1104CA: Instructions for Fault Alarm and Lifespan
FDA (Devices) · 2023-12-13
- HighMedical Device Instruction Update: HVAD Ventricular Assist Device Clarifications
FDA (Devices) · 2023-12-13
- HighHVAD Ventricular Assist Device Updated Instructions for Alarm Conditions
FDA (Devices) · 2023-12-13
Same hazard · material hardening
See all →- HighHeartWare HVAD Pump Kit driveline cover may harden, impeding urgent access
FDA (Devices) · 2022-12-28
- HighHeartWare HVAD Pump Kit Driveline Cover Hardening Issue
FDA (Devices) · 2022-12-28
Similar recalls
- CriticalHeartWare Ventricular Assist System Recalled for Driveline Connector Pin Retraction
FDA (Devices) · 2015-06-17
- CriticalHeartWare HVAD Driveline Extension Cable Recalled for Cardiac Safety Concerns
FDA (Devices) · 2021-08-18
- SevereHeartware Recalls HVAD Pump Implant Kits Due to Driveline Connector Loosening
FDA (Devices) · 2013-05-22
- HighMedtronic HVAD Pump Implants: Driveline Cover May Impede Connector Access
FDA (Devices) · 2021-09-22
- SevereHeartWare HVAD Pump Recalled for Foreign Material in Driveline Connector
FDA (Devices) · 2016-10-19
- CriticalHeartWare HVAD System Recalled Due to Higher Mortality Rates
FDA (Devices) · 2021-08-18