The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0685-2013·Announced 2013-01-23

DePuy Spine EXPEDIUM Offset Torque Wrench Recalled for Wear

DePuy Spine is recalling 140 EXPEDIUM Offset Torque Wrenches because a handle component wears prematurely, potentially causing inaccurate torque readings.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device where a defect could lead to inaccurate torque readings during a surgical procedure. This represents a risk-of-harm product where injury has not yet been reported, fitting the High severity criteria.

Plain-English summary

DePuy Spine, Inc. is recalling 140 units of the EXPEDIUM Offset Torque Wrench (Product Code 2797-29-510). The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.

The recall is due to a component within the handle that is prematurely wearing. This wear may result in the instrument recording torque values outside of the specification.

The affected units were distributed nationwide in the United States and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK. All lots are affected. No specific instructions for consumers are provided in the source text.

The recalled product

Product
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Manufacturer
DePuy Spine, Inc.
Affected units
140

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All lots

Distribution

Distributed nationwide across the United States.