Biomet Recalls PT Hybrid Glenoid Post Due to Oversized Thread
Biomet is recalling PT Hybrid Glenoid Posts because an oversized male thread may prevent the post from seating fully, potentially creating a gap in the implant.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a structural defect in a medical device that could lead to implant failure or complications, but the source text does not report any injuries, illnesses, or deaths, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Biomet, Inc. is recalling the PT113950 PT Hybrid Glenoid Post, a sterile, semi-constrained shoulder prosthesis component. The recall follows an investigation that identified the male thread of the post may be oversized. This condition can vary in degree and may lead to the post not being fully seated into the Hybrid Base if the implant construct is implanted, resulting in a gap of 1-3 mm.
The affected product is distributed worldwide, including the United States, Canada, Japan, Argentina, Australia, the Netherlands, Europe, and Chile. The recall covers 1,050 units across multiple specific lot numbers.
Consumers and healthcare providers should contact Biomet for further instructions regarding the return or replacement of these specific lots.
The recalled product
- Product
- PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Manufacturer
- Biomet, Inc.
- Affected units
- 1,050
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Part 113950 Lot 093540
- 093550
- 093570
- 093580
- 093590
- 093600
- 093610
- 093620
- 093720
- 093730
- 093740
- 093760
- 093770
- 093780
- 282590
- 282600
- 282620
- 282630
- 282680
- 282700
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Biomet, Inc.
See all →- ModerateIntramedullary Bone Saw Blade Assembly Recalled Due to Potential Packaging Damage
FDA (Devices) · 2026-08-12
- HighIntramedullary Bone Saw Blade Assemblies Recalled Due to Potential Packaging Damage
FDA (Devices) · 2026-08-12
- ModerateIntramedullary Bone Saw Blade Assembly Recalled Due to Potential Sterile Barrier Damage
FDA (Devices) · 2026-08-12
- HighIntramedullary Bone Saw Blade Assembly Recalled Due to Potential Sterile Barrier Damage
FDA (Devices) · 2026-08-12
- HighIntramedullary Bone Saw Blade Assembly Recall for Packaging Damage
FDA (Devices) · 2026-08-12
Same hazard · structural defect
See all →- SevereSubaru Recalls 2026 Forester and Crosstrek for Ball Joint Defect
NHTSA · 2026-10-19
- SevereVolvo Recalls 2025-2026 VNL Trucks for Loose Structural Bolts
NHTSA · 2026-10-15
- Severe
- SevereVolvo Recalls EX30 and EX30CC Vehicles for Head Restraint Defect
NHTSA · 2026-10-09
- Severe
Similar recalls
- HighBiomet Recalls Shoulder Implant Components Due to Endotoxin Levels
FDA (Devices) · 2020-06-03
- ModerateZimmer Biomet Recalls Modular Hybrid Glenoid Base for Quality Issues
FDA (Devices) · 2017-06-07
- HighBiomet Shoulder Surgery Pliers Recalled for Sterility Concerns
FDA (Devices) · 2020-04-29
- LowBiomet Shoulder Joint Replacement Components Recalled for Incorrect GTIN Codes
FDA (Devices) · 2015-06-17
- HighBiomet Recalls Offset Tibial Tray Adaptors Due to Sizing Issue
FDA (Devices) · 2012-12-26
- HighBiomet Shoulder Implant Stem Coating Adhesion Defect Recall
FDA (Devices) · 2022-03-02