Deerfield Imaging is recalling 14 IMRIS ORT 300 operating room tables because they may drift in the roll position while in use.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a mechanical defect that could affect patient safety during procedures, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Deerfield Imaging, Inc. is recalling 14 units of the IMRIS ORT 300 Removable Operating Room Table. The affected units include part numbers 114093-000 and 114093-006. These tables are intended for use during diagnostic examinations, radiologic procedures, or surgical procedures to support and position a patient.

The recall is being conducted because the table may drift in the roll position while in use. This defect could potentially compromise patient positioning during procedures.

The affected units were distributed worldwide, including the United States, China, Qatar, Canada, Australia, Japan, France, Germany, Sweden, and Italy. Healthcare facilities should contact Deerfield Imaging for instructions on how to address the issue.

The recalled product

Product
IMRIS ORT 300, Removable Operating Room Table, Part Numbers: (a) 114093-000, (b) 114093-006 - Product Usage: The ORT200 and ORT300 are intended for use during diagnostic examinations, radiologic procedures, or surgical procedures to support and position a patient. The tables are
Affected units
14

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (14)

  • (a) 114093-000
  • Serial Numbers: 10005695
  • 10002998
  • 10005456
  • 10007377
  • 10002997
  • 10005612
  • 10004372
  • 10002841 10002255
  • 10002385
  • 10005613 (b) 114093-006
  • Serial Numbers: 10007411
  • 10002996
  • 10004515

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 4 recalled by Deerfield Imaging, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 4 products in this recall →