OrthoPediatrics Recalls Response Spine Pedicle Screws Due to Manufacturing Defect
OrthoPediatrics Corp is recalling 102 Response Spine UniAxial Pedicle Screws because they may have been manufactured incorrectly, preventing the driver from fully engaging the screw.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the defect in a spinal implant represents a risk-of-harm product where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
OrthoPediatrics Corp is recalling 102 units of the Response 5.5/6.0 Spine System, specifically the UniAxial Pedicle Screw and UniAxial Reduction Pedicle Screw. The recall was initiated because the devices may have been manufactured incorrectly, which prevents the driver from fully engaging into the hexagonal drive feature of the screw.
The affected products were distributed in the United States to the states of Alabama, Arizona, California, Delaware, Florida, Georgia, Hawaii, Kentucky, Louisiana, Michigan, Missouri, Minnesota, Mississippi, North Carolina, New Jersey, New York, Ohio, Oregon, Tennessee, and Texas. The specific item numbers, lot numbers, and product descriptions for the affected units are listed in the structured product information.
Healthcare providers and patients should stop using the affected devices and contact OrthoPediatrics Corp for further instructions on how to proceed with the recall.
The recalled product
- Product
- UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative…
- Manufacturer
- OrthoPediatrics Corp
- Affected units
- 102
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Item number
- lot
- product description 00-1300-1735
- 44003
- 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 35MM. 00-1300-0735
- 44601
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 35MM. 00-1300-0740
- 44602
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740
- 43926
- 44252
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0745
- 43927
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 45MM. 00-1300-0750
- 44619
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 50MM. 00-1300-0620
- 44623
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 20MM. 00-1300-0640
- 44617
- 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-0640
Distribution
Related recalls
Same manufacturer · OrthoPediatrics Corp
See all →- HighOrthopedic Implant Bolts Recalled Due to Breakage Risk in Bone Correction
FDA (Devices) · 2023-02-22
- HighExternal Fixation Wire May Fracture During Surgery, Leaving Fragments in Patient
FDA (Devices) · 2022-09-14
- SevereOrthoPediatrics Recalls Spine Rods Due to Potential Fracture Risk
FDA (Devices) · 2020-08-05
- ModerateOrthoPediatrics Recalls ShieldLoc ACL Implant Components Due to Labeling Errors
FDA (Devices) · 2020-07-15
- ModerateOrthoPediatrics Recalls ShieldLoc ACL Implant Components Due to Labeling Errors
FDA (Devices) · 2020-07-15
Same hazard · manufacturing defect
See all →- HighBeak & Skiff Alcohol Antiseptic Sanitizer recalled for acetaldehyde contamination
FDA (Drugs) · 2026-10-14
- HighFDA Recalls Opportunity Pro Nitrile Gloves Due to Manufacturing Defects
FDA (Devices) · 2026-10-07
- HighWright Medical Recalls Charlotte Claw Screws Due to Missing Threads
FDA (Devices) · 2026-10-07
- ModerateCustom Interventional Pain Management Recalls Mastisol Skin Adhesive Trays
FDA (Devices) · 2026-10-07
- HighNOxBOXi Nitric Oxide Delivery System Recalled for Oxygen Sensor Defect
FDA (Devices) · 2026-10-07
Similar recalls
- SevereOrthoPediatrics Recalls Spine Rods Due to Potential Fracture Risk
FDA (Devices) · 2020-08-05
- ModerateOrthoPediatrics Recalls 7.0mm Pedicle Screws Due to Color Anodizing Error
FDA (Devices) · 2018-02-21
- ModerateSpineFrontier Recalls PedFuse Reset Pedicle Screws Due to Assembly Error
FDA (Devices) · 2013-12-25
- HighOrthofix Recalls Connector System Set Screws Due to Size Discrepancies
FDA (Devices) · 2018-04-18
- HighAlphatec Recalls Leucadia Autolok Pedicle Screw System Due to Tolerancing Issue
FDA (Devices) · 2013-06-05
- ModerateOrthoPediatrics Recalls 3.5mm Locking Cortical Bone Screws Due to Packaging Error
FDA (Devices) · 2014-02-12