The Recall Desk
HighFDA (Devices)·Z-0620-2025·Announced 2024-12-11

Olympus Removes Acecide-C Disinfection from GIF-1TH190 Endoscope Reprocessing

Olympus has removed high-level disinfection with Acecide-C as an approved reprocessing method for the GIF-1TH190 gastrointestinal videoscope. Approximately 1,770 units nationwide are affected.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: FDA Class II removal of an approved high-level disinfection method for an internal gastrointestinal endoscope. No reported illnesses or injuries, but inadequate sterilization of internal devices poses infection risk. Fits rubric criterion: risk-of-harm products where injury has not yet been reported.

Plain-English summary

Olympus Corporation of the Americas has removed high-level disinfection (HLD) with Acecide-C disinfectant from the approved reprocessing methods for the EVIS EXERA III Gastrointestinal Videoscope (Model GIF-1TH190). The Instructions for Use for the Olympus OER-Pro Automated Endoscope Reprocessor has been updated to reflect this change.

Approximately 1,770 units of the GIF-1TH190 have been distributed nationwide across all U.S. states, with serial numbers before 2446103 or after 2500453 identified as affected.

Acecide-C HLD is no longer an approved method for reprocessing these endoscopes. Facilities and healthcare providers should consult the updated Instructions for Use for the Olympus OER-Pro Automated Endoscope Reprocessor to identify approved reprocessing methods.

The recalled product

Product
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
Manufacturer
Olympus Corporation of the Americas
Hazard
  • inadequate-disinfection
  • infection-risk

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (4)

  • Model Number: GIF-1TH190
  • UDI: 04953170343360
  • 04953170437014
  • Serial Number Range: Before 2446103 or After 2500453

Distribution

Distributed nationwide across the United States.