Zimmer Spine Ardis Inserter Recalled for Potential Spacer Fragmentation
Zimmer, Inc. is recalling the Spine Ardis Inserter because excessive force can break the PEEK spacer, causing sharp fragments that may damage adjacent structures.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up…"
Plain-English summary
Zimmer, Inc. is recalling the Zimmer Spine Ardis Inserter, a prescription, non-sterile medical device. The recall involves 315 units, with 172 distributed in the United States and the remainder distributed worldwide in Australia, El Salvador, France, and Singapore.
The recall was initiated after Zimmer Spine received reports that the PEEK Ardis Interbody Spacer can break into fragments if too much lateral force is applied to the Ardis Inserter. If the spacer breaks during implantation, it can cause surgical delays of up to 60 minutes. Additionally, the sharp edges of the broken implant may contact and damage adjacent structures, potentially requiring further surgical intervention.
Healthcare providers and patients should contact Zimmer, Inc. for instructions on how to handle the recalled devices. The source text does not specify a specific action for consumers, but the device is a prescription item used in surgical settings.
The recalled product
- Product
- Zimmer Spine Ardis Inserter, Rx, non-sterile
- Manufacturer
- Zimmer, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Ardis Inserter
- Part number: 3256-01
- Lot: 52ZW
- 50CL
- 53BB
- 55AB
- 58GS
- 58GT
- 60JY
- 60JZ
- 62KP
- 63HJ
- 64EY
- 68AX
- 71JE
- 75LE Part
- 07.00780.701
- lot 76CN
- part 07.00780.702
- lot 76CP
Distribution
Related recalls
Same manufacturer · Zimmer, Inc.
See all →- HighZimmer Off-Axis Alliance Glenoid Reamer Guide Plastic Breakage Recall
FDA (Devices) · 2026-07-01
- HighZimmer Off-Axis Alliance Glenoid Reamer Guide recalled for breakage
FDA (Devices) · 2026-07-01
- HighZimmer Shoulder Reamer Guide System Recalled for Plastic Breakage
FDA (Devices) · 2026-07-01
- ModeratePersona Revision Trabecular Metal Femoral Distal Augment recalled for packaging defect
FDA (Devices) · 2026-07-01
- HighZimmer Off-Axis Comprehensive Shoulder System Reamer Guide Recall
FDA (Devices) · 2026-07-01
Same hazard · structural failure
See all →- High
- ModerateQuantum Ops Recalls ADAPTABLE Beach Chair Surgical Patient Positioners
FDA (Devices) · 2026-10-14
- SevereMedline Bariatric Rollator Recalled Due to Front Wheel Assembly Disconnect Risk
FDA (Devices) · 2026-10-14
- SevereMedline Bariatric Rollator Recalled Due to Front Wheel Detachment Risk
FDA (Devices) · 2026-10-14
- SeverePacearth Recalls Swing Seats Due to Serious Injury Risk from Falling
CPSC · 2026-10-08
Similar recalls
- SevereZimmer Ardis Interbody System Inserter Recalled for Release Knob Issues
FDA (Devices) · 2013-01-16
- ModerateZimmer Segmental Systems Polyethylene Inserts Recalled for Loading Instructions
FDA (Devices) · 2014-01-29
- HighZimmer Persona Tibial Inserter Recalled for Potential Tip Fracture
FDA (Devices) · 2015-02-11
- HighZimmer Biomet Recalls Spinal Rod Cutters Due to Fracture Risk
FDA (Devices) · 2020-12-30
- HighZimmer Biomet Recalls Persona Partial Knee System Impactor Pads
FDA (Devices) · 2018-02-21
- HighZimmer Biomet Recalls Persona Partial Knee Spacer Blocks
FDA (Devices) · 2018-04-04