Zimmer Spine Ardis Inserter Recalled for Potential Spacer Fragmentation
Zimmer, Inc. is recalling the Spine Ardis Inserter because excessive force can break the PEEK spacer, causing sharp fragments that may damage adjacent structures.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product where injury (damage to adjacent structures) is a potential consequence, but no specific injuries or illnesses are reported in the source text, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Zimmer, Inc. is recalling the Zimmer Spine Ardis Inserter, a prescription, non-sterile medical device. The recall involves 315 units, with 172 distributed in the United States and the remainder distributed worldwide in Australia, El Salvador, France, and Singapore.
The recall was initiated after Zimmer Spine received reports that the PEEK Ardis Interbody Spacer can break into fragments if too much lateral force is applied to the Ardis Inserter. If the spacer breaks during implantation, it can cause surgical delays of up to 60 minutes. Additionally, the sharp edges of the broken implant may contact and damage adjacent structures, potentially requiring further surgical intervention.
Healthcare providers and patients should contact Zimmer, Inc. for instructions on how to handle the recalled devices. The source text does not specify a specific action for consumers, but the device is a prescription item used in surgical settings.
The recalled product
- Product
- Zimmer Spine Ardis Inserter, Rx, non-sterile
- Manufacturer
- Zimmer, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Ardis Inserter
- Part number: 3256-01
- Lot: 52ZW
- 50CL
- 53BB
- 55AB
- 58GS
- 58GT
- 60JY
- 60JZ
- 62KP
- 63HJ
- 64EY
- 68AX
- 71JE
- 75LE Part
- 07.00780.701
- lot 76CN
- part 07.00780.702
- lot 76CP
Distribution
Related recalls
Same manufacturer · Zimmer, Inc.
See all →- HighZimmer Off-Axis Alliance Glenoid Reamer Guide recalled for breakage
FDA (Devices) · 2026-07-01
- ModerateZimmer Shoulder System Reamer Guide recalled for plastic breakage
FDA (Devices) · 2026-07-01
- ModeratePersona Revision Trabecular Metal Femoral Distal Augment recalled for packaging defect
FDA (Devices) · 2026-07-01
- HighZimmer Shoulder Reamer Guide System Recalled for Plastic Breakage
FDA (Devices) · 2026-07-01
- HighZimmer Off-Axis Alliance Glenoid Reamer Guide Plastic Breakage Recall
FDA (Devices) · 2026-07-01
Same hazard · structural failure
See all →- SevereMondraker Recalls Mountain Bicycles with ONOFF S9 Carbon Handlebars
CPSC · 2026-09-03
- HighFord F-650 SD Recall: Liftgate Platform Deck Pins May Break
NHTSA · 2026-08-20
- HighStryker Hoffmann II Connecting Rod Micro May Crack After Cleaning and Sterilization
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Connecting Rod Micro May Crack After Cleaning and Sterilization
FDA (Devices) · 2026-08-05
- High