Siemens Recalls Advia Chemistry Calibrator Due to Bilirubin Bias
Siemens is recalling Advia Chemistry Calibrators because incorrect values may cause a 41% positive bias in Direct Bilirubin patient results.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device defect that results in significant diagnostic inaccuracy (41% bias), which poses a risk of harm through incorrect clinical decisions, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 10,614 units of Advia Chemistry Calibrators. The affected lots are 680726, 680726A, 680726B, 680726C, and 680726D. These devices are used for in vitro diagnostic calibration of ADVIA Chemistry systems.
The recall was initiated because the Assigned Calibrator Values (ACV) for the Direct Bilirubin (DBIL_2) method may cause a positive proportional bias of approximately 41% on patient samples compared to the reference method. This error could lead to inaccurate patient test results.
The affected products were distributed nationwide in the United States and internationally to numerous countries including Canada, Mexico, Australia, and various nations in Europe, Asia, and South America. Healthcare facilities and laboratories should stop using the affected calibrators and contact Siemens for instructions on replacement or return.
The recalled product
- Product
- Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CRE
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 10,614
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Device Listing # D033650 Model# SMN# 10312279 Lot Numbers: 680726
- 680726A
- 680726B
- 680726C
- 680726D
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- ModerateSiemens Tricyclic Antidepressant Calibrator/Controls Recall Lot U2
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · mis labeling
See all →- ModerateMedline Trach Tube Kit Recalled for Component Labeling Discrepancy
FDA (Devices) · 2026-09-02
- ModerateAVID Medical Recalls Halyard Total Joint Kits for Latex Labeling Deficiency
FDA (Devices) · 2026-09-02
- ModerateOlympus Recalls PowerSeal Sealer Due to Incorrect Outer Box Labeling
FDA (Devices) · 2026-09-02
- HighSura Tanmen Ramen Noodles Recalled for Undeclared Fish Allergens
FDA (Food) · 2026-09-02
- SevereWhite Vein Advanced Alkaloids Chewable Tablets Recall for Excess Active Ingredient
FDA (Food) · 2026-09-02