Inpeco Laboratory Automation Systems Recalled for Burnt Sockets
Inpeco is recalling FlexLab, Accelerator a3600, and Aptio Automation systems because field reports indicate burnt Schuko sockets and smoke.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (electrical fire/smoke) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Inpeco S.A. is recalling 188 units of FlexLab, Accelerator a3600, and Aptio Automation modular laboratory systems. These systems are designed to automate pre-analytical and post-analytical processing and sample handling in laboratory settings.
The recall was initiated after Inpeco received several notifications from the field regarding burnt Schuko sockets. Customers reported seeing smoke coming from the damaged sockets. The affected components include specific Schuko sockets with part numbers 693822160, 693822170, 693822190, and 0C00005640, associated with the CMSA/HETTICH - STORAGE - i2000SR model.
The affected units were distributed worldwide, including in the United States (specifically Texas and New York) and foreign consignees in Germany, Italy, France, Austria, and Spain. Users should identify their equipment by checking the serial numbers listed in the official recall documentation.
The recalled product
- Product
- FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a…
- Manufacturer
- Inpeco S.A.
- Hazard
- Affected units
- 188
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Shucko Socket with P/N: 693822160
- Model: CMSA/HETTICH - STORAGE - i2000SR
- Length: 3 meters
- Shucko Socket with P/N: 693822170
- Length: 5 meters
- Shucko Socket with P/N: 693822190
- Length: 7 meters
- Shucko Socket with P/N: 0C00005640
- ACP.024
- ACP.039
- ACP.054
- ACP.069
- ACP.084
- ACP.100
- ACP.115
- ACP.010
- ACP.025
- ACP.040
- ACP.055
- ACP.070
Distribution
Related recalls
Same manufacturer · Inpeco S.A.
See all →- HighLaboratory automation recalled over unlabelled or mismatched secondary sample tubes
FDA (Devices) · 2025-12-03
- HighLaboratory automation recalled over sample tubes with mismatched labels
FDA (Devices) · 2025-12-03
- HighLaboratory automation systems recalled over unevaluated sample integrity function
FDA (Devices) · 2025-10-15
- HighFlexLab X system recalled over unevaluated blood sample integrity function
FDA (Devices) · 2025-10-15
- ModerateRoche Cobas Interface Module Firmware May Cause Delayed Sample Processing
FDA (Devices) · 2023-12-27
Same hazard · fire
See all →- SevereVQQQ Gas Patio Heaters Recalled for Fire and Burn Hazards
CPSC · 2026-10-22
- HighForest River Recalls 2023-2026 Pause Travel Trailers for Wiring Defect
NHTSA · 2026-10-20
- SevereForest River Recalls 2026-2027 Cedar Creek Trailers for Fire Risk
NHTSA · 2026-10-16
- SevereAlliance RV Recalls Valor and Delta Trailers Due to Fire Risk
NHTSA · 2026-10-16
- SevereSpartan RV Chassis K2 Recalled for Fuel Hose Separation Risk
NHTSA · 2026-10-15
Similar recalls
- HighInpeco Laboratory Automation Systems Recalled for Module Freezing Risk
FDA (Devices) · 2018-05-16
- HighInpeco FlexLab and Accelerator a3600 Recalled for Missing Safety Labels
FDA (Devices) · 2018-11-07
- HighInpeco FlexLab Automation System Recalled for Potential Sample Dilution
FDA (Devices) · 2020-11-18
- HighInpeco Aptio Automation System Recall for Potential Sample Dilution
FDA (Devices) · 2020-11-18
- HighInpeco Accelerator a3600 Aliquoter Module Recall for Sample Dilution Risk
FDA (Devices) · 2020-11-18
- HighInpeco FlexLab 3.6 Automation Recalled for Potential Sample ID Mis-association
FDA (Devices) · 2020-09-23