Sensica Urine Output System displays incorrect timestamps when connected to WiFi
The Sensica Urine Output System (SCCS1002) displays incorrect timestamps and monitoring data when connected to WiFi or internet due to a time synchronization issue. This malfunction affects the accuracy of patient monitoring records.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This FDA Class II medical device recall involves data integrity failure in a patient monitoring system. Although no deaths, injuries, or illnesses are reported, incorrect timestamp data in a urine output monitoring device constitutes a risk-of-harm product where inaccuracy could impact clinical decision-making. Per the rubric, when injuries are not reported the maximum score is 3.
Plain-English summary
The Sensica Urine Output System, Catalog Number SCCS1002, manufactured by C.R. Bard Inc, is being recalled due to a WiFi connectivity issue that causes incorrect time display and data recording.
When the device is connected to WiFi or internet, the displayed time may change during a patient monitoring session. This causes incorrect times to appear on the monitoring screen, incorrect timestamps in hour block data, incorrect timestamps in data screens, and incorrect catheter use time records.
The recall affects 148 units distributed nationwide in the United States. Affected units are identified by serial numbers listed under UDI/DI 00801741215773. Healthcare providers using affected devices should contact C.R. Bard Inc for guidance on corrective measures.
The recalled product
- Product
- Sensica Urine Output System, Catalog Number SCCS1002
- Manufacturer
- C.R. Bard Inc
- Affected units
- 148
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI/DI 00801741215773
- Serial numbers: BMGPRMYS001
- BMGPRMYS003
- BMGPRMYS004
- BMGPRMYS005
- BMGPRMYS006
- BMGPRMYS007
- BMGPRMYS020
- BMGPRMYS026
- BMGPRMYS028
- BMGPRMYS030
- BMFWRMSBT4
- BMFWRMSJ45
- BMFWRMSJ49
- BMFZRMS104
- BMGNRMS085
- BMFSRMSH98
- BMFSRMSJ01
- BMFVRMSQ62
- BMFVRMSQ66
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · C.R. Bard Inc
See all →- HighC.R. Bard Recalls SureStep Foley Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- HighC.R. Bard Recalls Foley Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- ModerateC.R. Bard Recalls Foley Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- HighC.R. Bard Recalls SureStep Foley Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- HighC.R. Bard Recalls SureStep Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
Similar recalls
- HighSensica Urine Output System recalled for measurement accuracy complaints
FDA (Devices) · 2022-07-06
- HighC.R. Bard Recalls CritiCore Automated Urine Output and Temperature Monitor
FDA (Devices) · 2018-06-27
- HighC.R. Bard Recalls CritiCore Monitors Due to Measurement and Alarm Failures
FDA (Devices) · 2018-02-21
- HighC.R. Bard Recalls Uros Automated Urine Output and Temperature Monitor
FDA (Devices) · 2018-02-07
- HighC.R. Bard Recalls SureStep Catheter Trays Due to Wipe Contamination
FDA (Devices) · 2026-09-16
- ModerateCapso Vision Recalls CapsoCam Plus Capsule Endoscopy Systems Due to Labeling Errors
FDA (Devices) · 2020-05-27