ImagePilot diagnostic imaging software versions 1.92 and 1.93 measurement error recall
Konica Minolta's ImagePilot versions 1.92 and 1.93 software may provide incorrect measurement calculations if improper power sequences are used with AeroDR or Momentum Panel systems.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This FDA Class II recall involves a risk-of-harm product where measurement inaccuracy in diagnostic imaging could impact patient care. No illnesses or injuries have been reported, and the hazard is contingent on improper power sequencing.
Plain-English summary
Konica Minolta Healthcare Americas, Inc. is recalling ImagePilot versions 1.92 and 1.93, medical imaging software designed to process and present medical images on display monitors for diagnostic purposes. When used with AeroDR or Momentum Panel imaging systems, these software versions can produce incorrect length measurements or area calculations if an incorrect power sequence is followed.
The software is used in diagnostic imaging facilities across the United States. Approximately 30 systems containing the affected software versions have been distributed nationwide. The company has identified that improper power sequencing during operation can lead to inaccurate diagnostic measurements.
Healthcare providers and clinicians using these software versions should verify their power sequence procedures match the manufacturer's specifications. Users should contact Konica Minolta Healthcare Americas, Inc. for guidance on identifying if their systems are affected and for information on obtaining corrected software versions.
The recalled product
- Product
- ImagePilot Versions 1.92 and 1.93-software primarily facilitates processing and presentation of medical images on display monitors for medical diagnostics Product number: ADFR and ADJC
- Manufacturer
- Konica Minolta Healthcare Americas, Inc.
- Hazard
- Affected units
- 30
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Konica Minolta Healthcare Americas, Inc.
See all →- HighKonica Minolta Recalls 151 X-Ray Systems Due to Unintentional Movement
FDA (Devices) · 2018-12-26
Same hazard · measurement error
See all →- ModeratePhilips MR 7700 MRE Upgrade MRI Equipment Stiffness Value Error Risk
FDA (Devices) · 2026-05-06
- HighPhilips Evolution Upgrade 1.5T MRI Stiffness Calculation Error
FDA (Devices) · 2026-05-06
- ModeratePhilips SmartPath to Ingenia Elition X MR Elastography Stiffness Value Errors
FDA (Devices) · 2026-05-06
- ModeratePhilips Ingenia Ambition X MRI Systems Stiffness Value Error Recall
FDA (Devices) · 2026-05-06
- ModeratePhilips MR 7700 MR Elastography device stiffness value error recall
FDA (Devices) · 2026-05-06
Similar recalls
- HighMIM Radiological Imaging Software Versions May Generate Incorrect Diagnostic Values
FDA (Devices) · 2025-03-12
- HighArietta 750 Ultrasound Software Versions Affected by Incorrect Measurement Display
FDA (Devices) · 2021-11-10
- HighFUJIFILM Surpria Imaging Software May Fail During Scanning
FDA (Devices) · 2022-09-21
- HighMedical Device Imaging Software May Calculate Incorrect Breast Measurements
FDA (Devices) · 2024-03-13
- HighMedical imaging system software errors may delay treatment procedures
FDA (Devices) · 2022-04-27
- HighKonica Minolta Sonimage HS1 Ultrasound System Recalled for Inaccurate Doppler Measurements
FDA (Devices) · 2017-08-02