Blood Ketone Test Strips Recalled for False Negative Results
Stanbio Laboratory is recalling STAT-Site M BHB Test Strips due to stability deterioration that causes false negative results. This defect may delay diagnosis and treatment of patients with conditions such as diabetic ketoacidosis.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall with no reported illnesses or injuries. However, the defect creates significant risk by producing false negative results that could delay critical diagnosis and treatment of patients with serious conditions such as diabetic ketoacidosis.
Plain-English summary
Stanbio Laboratory is recalling 2,351 units of STAT-Site M BHB Test Strips, which are used with the STAT-Site M Beta-HB Meter to measure blood ketone values. The strips are distributed nationwide in the US and also in Canada and Singapore.
The test strips are affected by stability deterioration caused by a pH shift in the coating solution during manufacturing. This instability reduces the strips' ability to detect low-range ketone values and causes them to fail to meet promised linearity performance specifications.
As a result of these defects, the test strips may produce false negative results—indicating lower or absent ketones when they are actually present. This can lead to delayed diagnosis and treatment initiation for patients with conditions such as diabetic ketoacidosis.
The recalled product
- Product
- STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.
- Manufacturer
- Stanbio Laboratory, LP
- Affected units
- 2,351
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- UDI-DI: 00657498000090
- Lot/Expiration: 203044/2024-01-31
- 209644/2024-06-30
Distribution
Related recalls
Same manufacturer · Stanbio Laboratory, LP
See all →- HighBreast Milk Analyzer Lacks FDA Validation, Risks Sub-optimal Infant Nutrition
FDA (Devices) · 2022-04-20
- ModerateBeta-Hydroxybutyrate Diagnostic Reagent Recalled for Stability Degradation
FDA (Devices) · 2021-08-11
- ModerateBeta-Hydroxybutyrate diagnostic reagent recalled due to stability degradation
FDA (Devices) · 2021-08-11
- ModerateIn Vitro Diagnostic Reagent Recalled Due to Stability Deterioration
FDA (Devices) · 2021-08-11
- HighIn Vitro Diagnostic Reagent Recalled for Reagent Stability Deterioration
FDA (Devices) · 2021-08-11
Same hazard · false negatives
See all →- HighSiemens Atellica Microalbumin Diagnostic Test Kits Recalled for Measurement Error
FDA (Devices) · 2024-09-25
- HighBioFire FilmArray GI Panel Recalled for Potential False Negative Results
FDA (Devices) · 2023-04-26
- HighBlood Culture Identification Panel Recalled Due to Manufacturing Defect
FDA (Devices) · 2021-06-30
- HighFilmArray Pneumonia Panel diagnostic test kits recalled for false results
FDA (Devices) · 2021-06-30
Similar recalls
- HighMedline Stat-Site B-HB Diluent Recalled for Potential Test Inaccuracy
FDA (Devices) · 2021-04-21
- SevereBlood Glucose Ketone Meter Firmware Bug Causes Inaccurate Test Results
FDA (Devices) · 2025-01-01
- HighPolymedco OC-Light iFOBT Test Strips Recalled for Potential False Negatives
FDA (Devices) · 2013-04-10
- HighPolymedco Recalls OC-Light iFOBT Test Strips Due to Stability Issues
FDA (Devices) · 2013-04-10
- LowMedline Recalls STAT-Site B-HB Controls Due to Shipment Compromise
FDA (Devices) · 2021-03-03
- SevereBlood glucose ketone meter software bug causes erroneous result transmission
FDA (Devices) · 2025-01-01