Luminex Recalls Verigene Processor SP Due to Unapproved Service Fixtures
Luminex is recalling 2,838 Verigene Processor SP units because field service engineers used unapproved test fixtures to calibrate and verify the devices.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves the use of unapproved and unvalidated test fixtures for calibration, which is a quality control issue. No specific injuries, illnesses, or deaths are reported in the source text, and the defect is related to service equipment rather than a direct structural failure or contamination of the device itself.
Plain-English summary
Luminex Corporation is recalling 2,838 units of the Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. These devices are used with the VERIGENE Gram-Positive Blood Culture Nucleic Acid Test, BC-GP, IVD. The recall covers units distributed worldwide, including the United States, Kuwait, Austria, the United Kingdom, Spain, France, Chile, Japan, the Netherlands, Greece, Italy, Thailand, Germany, and Turkey.
The recall was initiated because Luminex determined that three separate test fixtures used by field service engineers to service the processors were procured from unapproved sources. In some cases, these fixtures were manufactured using unvalidated processes. The fixtures were used to calibrate and verify the VERIGENE SP devices.
Healthcare facilities and laboratories that possess the affected Verigene Processor SP units should contact Luminex Corporation for instructions on how to proceed with the recall. Specific serial numbers for the affected units are listed in the official FDA recall record Z-0476-2021.
The recalled product
- Product
- Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigen
- Manufacturer
- Luminex Corporation
- Affected units
- 2,838
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI 00857573006010 Model Number 10-0000-07
- Serial Numbers: 10110051
- 10110050
- 10110048
- 10110047
- 10099033
- 10099032
- 10099029
- 10043027
- 10043025
- 10043024
- 10043023
- 10035106
- 10035105
- 10014003
- 10014001
- 9350165
- 9350164
- 9350163
- 9350161
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 2 recalled by Luminex Corporation under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 2 products in this recall →Related recalls
Same manufacturer · Luminex Corporation
See all →- HighLuminex VERIGENE Gram Negative Blood Culture Test Recalled
FDA (Devices) · 2025-06-11
- HighVerigene Gram Positive Blood Culture Test Recalled for Defective Cartridges
FDA (Devices) · 2024-08-14
- ModerateLuminex VERIGENE BC-GP Utility Trays Recalled Due to Potential Defects
FDA (Devices) · 2023-12-13
- HighVerigene nucleic acid test kits recalled for rare false-negative results
FDA (Devices) · 2023-07-05
- HighVERIGENE Clostridium difficile Nucleic Acid Test May Give False-Negative Results
FDA (Devices) · 2023-07-05
Same hazard · calibration error
See all →- ModeratePhotera 400 Phototherapy Device Recalled for Incorrect UV Sensor Calibration
FDA (Devices) · 2026-07-22
- ModeratePhotera 600 3-D phototherapy device UV sensor calibration error recall
FDA (Devices) · 2026-07-22
- ModeratePhotera 600 Phototherapy Device UV Sensor Calibration Error
FDA (Devices) · 2026-07-22
- HighPhothera 200 phototherapy calibration certificates had outputs reversed
FDA (Devices) · 2026-06-10
- HighRoche cobas analyzer may repeat previous results for spline calibrated assays
FDA (Devices) · 2026-03-11