Luminex Recalls Verigene Processor SP Due to Unapproved Service Fixtures
Luminex is recalling 2,838 Verigene Processor SP units because field service engineers used unapproved test fixtures to calibrate and verify the devices.
- Product
- Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigen
- Category
- Medical Device
- Distribution
- Distributed nationwide