The Recall Desk
ModerateFDA (Devices)·Z-0469-2018·Announced 2018-02-07

Zimmer Biomet Recalls Affixus Hip Fracture Nails Due to Assembly Defect

Zimmer Biomet is recalling 638 Affixus Hip Fracture Nails because a manufacturing rework error may have displaced internal components, potentially preventing proper assembly.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II. The hazard is a functional defect that prevents proper assembly, but the source text does not report any injuries, illnesses, or deaths, fitting the criteria for a moderate severity voluntary precautionary recall.

Plain-English summary

Zimmer Biomet, Inc. is recalling 638 units of the Affixus Hip Fracture Nail, RH 130 DEG 11MM X 400MM (Item No. 814511400). The affected units were distributed nationwide in the United States, as well as in Portugal and the Netherlands.

The recall was initiated after it was detected during a rework process for a previous field action that the set screw and sleeve assembly were no longer in their installed locations. These components are designed to have an interference fit, and their displacement could prevent the product from being properly assembled with the mating screw and targeting jig.

Healthcare providers and facilities that have received these specific units should stop using them and contact Zimmer Biomet for instructions on how to return or replace the affected devices.

The recalled product

Product
Affixus Hip Fracture Nail, RH 130 DEG 11MM X 400MM
Affected units
638

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Item No. 814511400

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 65 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.