GE Healthcare Recalls CARESCAPE R860 Inspiratory Safety Guards
GE Healthcare is recalling CARESCAPE R860 Inspiratory Safety Guards because they may disconnect from the breathing circuit, potentially causing loss of ventilation and hypoxia in patients.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class I, which mandates a severity score of at least 4 per the rubric. Additionally, the hazard involves a structural defect in a life-support device that could lead to hypoxia, a significant injury risk.
Plain-English summary
GE Healthcare, LLC is recalling the CARESCAPE R860 Inspiratory Safety Guard (part numbers 2066713-001 and 2083208-001). The recall involves 1,140 units worldwide, including 307 units in the United States. The affected lots are 17/00951, 17/01174, 17/01937, 17/01994, 17/02372, 17/02393, 18/00126, 18/00127, 18/00128, 18/00129, and 18/00130.
The recall was initiated because the Inspiratory Safety Guard may disconnect from the breathing circuit pathway. This disconnection could create a loss of ventilation, leading to inadequate oxygenation for patients and increasing the possibility of hypoxia. The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above.
The product was distributed worldwide, including the United States (nationwide including Puerto Rico) and numerous other countries. Healthcare facilities should contact GE Healthcare for instructions on how to address the issue and replace the affected components.
The recalled product
- Product
- GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b) 2083208-001 (box of ten singles) The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing…
- Manufacturer
- GE Healthcare, LLC
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (12)
- a) 2066713-001 Lot Numbers: 17/00951
- 17/01174
- 17/01937
- 17/01994
- 17/02372
- 17/02393
- 18/00126
- 18/00127
- 18/00128
- 18/00129
- 18/00130 b) 2083208-001 Lot Numbers: 17/00951
- 18/00130
Distribution
Related recalls
Same manufacturer · GE Healthcare, LLC
See all →- HighGE SIGNA Premier MRI magnets may have ferrous fittings fitted
FDA (Devices) · 2026-04-08
- HighGE Medical Imaging Software: Inaccurate Measurements in Global Stack Viewport
FDA (Devices) · 2022-08-31
- HighGE Healthcare CT, PET, and NM Patient Tables May Drop During Service
FDA (Devices) · 2022-08-17
- HighGE Centricity PACS Software fails to notify of exam study modifications
FDA (Devices) · 2022-08-03
- SevereGE Healthcare Anesthesia Systems Recalled for Potential Transducer Connection Reversal
FDA (Devices) · 2022-07-27
Same hazard · hypoxia
See all →- SevereCardinal Health Recalls Presource Packs with Medtronic Microstream Components
FDA (Devices) · 2026-10-14
- SevereGiraffe and Panda warmer blender knobs may deliver wrong oxygen concentration
FDA (Devices) · 2026-06-24
- SeverePanda freestanding warmer blender knobs may deliver wrong oxygen concentration
FDA (Devices) · 2026-06-24
- SevereGiraffe infant resuscitation system may deliver wrong oxygen concentration
FDA (Devices) · 2026-06-24
- HighOxygen masks recalled after complaints of tubing disconnecting during use
FDA (Devices) · 2026-01-14
Similar recalls
- HighGE Healthcare Recalls CARESCAPE R860 Ventilators Due to Oxygen Sensor Issue
FDA (Devices) · 2020-09-02
- SevereGE Healthcare Ventilators Recalled for Insufficient Battery Backup Power
FDA (Devices) · 2022-06-15
- HighDatex-Ohmeda CARESCAPE R860 Ventilators Recalled for Display Processor Failure
FDA (Devices) · 2018-06-27
- HighGE Healthcare Recalls CARESCAPE Respiratory Modules Due to O2 Sensor Failure
FDA (Devices) · 2014-09-17
- SevereVentilator Backup Battery Failure Recalled by GE Healthcare Worldwide
FDA (Devices) · 2022-06-15
- SevereDatex-Ohmeda Neonatal Flow Sensor Cable Recall for Missing Connector Housing
FDA (Devices) · 2018-08-15