Airway procedure kits recalled over cracks in anesthesia circuit tubing
Medline is recalling ENT and laryngoscopy-bronchoscopy kits containing anesthesia circuits after complaints of leaks from cracks in the 120-inch expandable tubing. The FDA classified it Class I.
What this means for you
Highest-severity recall. Risk of serious injury or death is documented or strongly suspected. Stop using the product now and follow the agency's instructions.
Our severity reasoning: Scored 5 because the FDA classified this recall as Class I, which the rubric assigns the top severity regardless of other factors.
Plain-English summary
Medline anesthesia circuit kits and other kits containing anesthesia circuits are being recalled, including DYNJ909128I (ENT) and DYNJ909114D (LARYNGOSCOPY - BRONCHOSCOPY).
The firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120 inch expandable tubing. Medline states that cracks and resulting leaks may lead to hypoxia, hypoventilation and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death, and that volatile anesthetics may also escape into the operating room environment.
The action covers 96,406 total units across the affected kits. The ENT kit carries case UDI-DI 40198459333355 and each UDI-DI 10198459333354, lot 25GBO685. Distribution was nationwide in the United States and internationally to Mexico and Panama. The FDA has classified the recall as Class I, its most serious category.
Facilities should quarantine affected lots immediately and follow Medline's instructions.
The recalled product
- Product
- Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
- Manufacturer
- Medline Industries, LP
- Category
- Medical Device — Surgical kit
- Affected units
- 96,406
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- ENT DYNJ909128I UDI-DI 40198459333355 (case)
- 10198459333354 (ea) Lots 25GBO685
- LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D UDI-DI 40198459055301 (case)
- 10198459055300 (ea) Lots 25GBN602
Distribution
Part of a larger recall action
This product is one of 10 recalled by Medline Industries, LP under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 10 products in this recall →Related recalls
Same manufacturer · Medline Industries, LP
See all →- HighMedline Recalls Cataract Preop Kits with Potentially Contaminated Applicators
FDA (Devices) · 2026-09-09
- SevereMedline C-Section Prep Kits Recalled for Potential Sterility Breach
FDA (Devices) · 2026-09-09
- HighMedline Blood Culture Kits Recalled Due to Potential Seal Defects
FDA (Devices) · 2026-09-09
- SevereMedline Eye and Cataract Kits Recalled for Potential Sterility Breach
FDA (Devices) · 2026-09-09
- SevereMedline Ultrasound Breast Pack Recalled for Potential Sterility Breach
FDA (Devices) · 2026-09-09
Same hazard · tubing crack
See all →- CriticalRobotic surgery kits recalled over cracked anesthesia circuit tubing
FDA (Devices) · 2025-11-19
- CriticalLaparoscopy and cysto kits recalled over cracked anesthesia circuit tubing
FDA (Devices) · 2025-11-19
- CriticalAdult anesthesia circuit kits recalled over cracked expandable tubing
FDA (Devices) · 2025-11-19
- CriticalSurgical convenience kits recalled over cracked anesthesia circuit tubing
FDA (Devices) · 2025-11-19
- CriticalOrthopedic and spine kits recalled over cracked anesthesia circuit tubing
FDA (Devices) · 2025-11-19