The Recall Desk

FDA (Devices) recall action 97882

Medline Industries, LP recalled 10 products on 2025-11-19

The FDA (Devices) publishes a recall like this one as 10 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 10 pages.

Products
10
Announced
2025-11-19
Critical
10
Units
964,060

Why these products were recalled

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–10 of 10

  • CriticalFDA (Devices)··2025-11-19

    Adult anesthesia circuit kits recalled over cracked expandable tubing

    Medline is recalling multiple adult 120-inch expandable anesthesia circuits after complaints of leaks caused by cracks in the tubing. The FDA has classified it Class I.

    Product
    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; Product description/Product numbers: ADULT 120 EXPANDABLE W/MASK DYNJAA10591; ANES CIRC 120IN EXP-LF DYNJAA6555G; ANES CIRCUIT ADULT 120"-LF DYNJAA10817; ANESTHESIA 120" ADLT-LF DYNJAA10
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Airway procedure kits recalled over cracks in anesthesia circuit tubing

    Medline is recalling ENT and laryngoscopy-bronchoscopy kits containing anesthesia circuits after complaints of leaks from cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; ENT DYNJ909128I; LARYNGOSCOPY - BRONCHOSCOPY DYNJ909114D
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Surgical convenience kits recalled over cracked anesthesia circuit tubing

    Medline is recalling 96,406 breast and hernia convenience kits after complaints of anesthesia circuit leaks caused by cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    Convenience Kit, BREAST HERNIA DYNJ909125G
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Robotic surgery kits recalled over cracked anesthesia circuit tubing

    Medline is recalling GYN Robotic and Robotic Prostate convenience kits after complaints of anesthesia circuit leaks caused by cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    Convenience kits; DYNJ901046M GYN ROBOTIC DYNJ909120D ROBOTIC PROSTATE
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Orthopedic and spine kits recalled over cracked anesthesia circuit tubing

    Medline is recalling ortho spine, total joint and kyphoplasty convenience kits after complaints of anesthesia circuit leaks from cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Breathing kits recalled over cracks in the expandable anesthesia tubing

    Medline is recalling the latex-free Breathing Kit DYNJ10550 after complaints of anesthesia circuit leaks caused by cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; BREATHING KIT-LF DYNJ10550
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Laparoscopy and cysto kits recalled over cracked anesthesia circuit tubing

    Medline is recalling robotic gastric sleeve, cysto and general laparoscopy kits after complaints of anesthesia circuit leaks from cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    DYNJ909119G ROBOTIC GASTRIC LAP SLEEVE-LF DYNJ909127F CYSTO DYNJ909129D GENERAL LAPAROSCOPY-LF
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Expandable anesthesia circuits recalled over cracks causing leaks in tubing

    Medline is recalling the 120-inch latex-free ExtendFlex anesthesia circuit after complaints of leaks caused by cracks in the expandable tubing. The FDA has classified it Class I.

    Product
    Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Urology and hip pinning kits recalled over cracked anesthesia circuit tubing

    Medline is recalling ortho spine, urology minor, set up and hip pinning convenience kits after complaints of anesthesia circuit leaks from cracked tubing. The FDA classified it Class I.

    Product
    Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
    Category
    Distribution
    34 states
  • CriticalFDA (Devices)··2025-11-19

    Open heart and vascular kits recalled over cracked anesthesia circuit tubing

    Medline is recalling open heart supplement and vascular kits containing anesthesia circuits after complaints of leaks from cracks in the 120-inch expandable tubing. The FDA classified it Class I.

    Product
    Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
    Category
    Distribution
    34 states