Intuitive Surgical Recalls 8mm Resano Forceps Due to Reprocessing Deviations
Intuitive Surgical is recalling 509 8mm Resano Forceps instruments because deviations in reprocessing steps can cause surface degradation and accelerated mechanical wear.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text does not report any hospitalizations, injuries, or illnesses, only potential mechanical wear and surface degradation.
Plain-English summary
Intuitive Surgical, Inc. is recalling 509 units of the 8mm Resano Forceps, intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments. The recall involves all 8mm da Vinci Xi (IS4000) Surgical System instruments manufactured with white colored housings, identified by part number 470181-04.
The recall was initiated because deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument. The affected devices were distributed in the United States to facilities in Florida, Texas, New York, Nevada, Oregon, Indiana, Utah, and Massachusetts.
Healthcare facilities should stop using the affected instruments and contact Intuitive Surgical for further instructions regarding the recall.
The recalled product
- Product
- 8MM RESANO FORCEPS,; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.
- Manufacturer
- Intuitive Surgical, Inc.
- Affected units
- 509
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Part number: 470181-04
Distribution
Part of a larger recall action
This product is one of 22 recalled by Intuitive Surgical, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 22 products in this recall →Related recalls
Same manufacturer · Intuitive Surgical, Inc.
See all →- Moderateda Vinci SP Instrument Arm Drape may detach during surgical procedures
FDA (Devices) · 2026-07-22
- SevereDa Vinci surgical system operative components with E-brake retainer pin defect
FDA (Devices) · 2026-07-01
- HighDa Vinci 5 tower may unintentionally activate the infrared handheld camera
FDA (Devices) · 2026-06-10
- Highda Vinci X and Xi Surgical Systems Recalled for Manipulator Arm Screw Breakage
FDA (Devices) · 2026-05-13
- SevereIntuitive 8 mm SureForm 30 Gray Reloads Incomplete Staple Defect
FDA (Devices) · 2026-05-13
Same hazard · mechanical wear
See all →- HighPhilips Azurion X-Ray Systems Table Movement Control Defect Recall
FDA (Devices) · 2026-05-27
- HighInfusion pumps recalled because degrading grease may cause excessive therapy delivery
FDA (Devices) · 2025-09-03
- HighX-ray C-arm Tabletop Recalled for Risk of Uncontrolled Motion
FDA (Devices) · 2024-05-15
- HighPhilips DigitalDiagnost C50 Radiographic System Ceiling Suspension Malfunction Risk
FDA (Devices) · 2023-04-19
- HighLiko Mobile Patient Lifts Recalled for Missing Bushings and Wear Risk
FDA (Devices) · 2022-02-23
Similar recalls
- HighSurgical Forceps Recalled for Frayed or Broken Grip Cables
FDA (Devices) · 2025-01-22
- HighIntuitive Surgical Recalls 8mm Precise Bipolar Forceps Due to Cautery Plug Defect
FDA (Devices) · 2013-06-05
- HighTenaculum Forceps for da Vinci Surgical Systems Recalled Due to Pitch Cable Failures
FDA (Devices) · 2025-01-22
- HighIntuitive Surgical Recalls 8mm Fenestrated Bipolar Forceps Due to Cautery Plug Defect
FDA (Devices) · 2013-06-05
- HighIntuitive Surgical Recalls 8MM Micro Bipolar Forceps Due to Frayed Grip Cables
FDA (Devices) · 2025-01-22
- HighIntuitive Surgical Recalls Fenestrated Bipolar Forceps for da Vinci System
FDA (Devices) · 2015-09-16