The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-0378-2027·Announced 2026-11-18

Vesco Q 1000mL Bag Feed Set Recalled for Inaccurate Fluid Delivery

Vesco Medical LLC is recalling 208,650 units of Vesco Q 1000mL Bag Feed Sets because the associated feeding pump may deliver enteral fluids outside its labeled accuracy specification.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as Class II, and the source describes a hazard where over- or under-delivery of enteral fluids could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Vesco Medical LLC is recalling 208,650 units of Vesco Q 1000mL Bag Feed Sets (PVC), Model Number VED-054. The recall covers all lots manufactured through 09/01/2026, identified by UDI-DI 00810096150639. The product was distributed nationwide in the United States.

The recall is being issued because the feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.

The source text does not provide specific instructions for consumers on how to dispose of or return the product, nor does it report any illnesses, injuries, or deaths associated with the use of these devices.

The recalled product

Product
Vesco Q 1000mL Bag Feed Set (PVC), Model Number VED-054
Manufacturer
Vesco Medical Llc
Affected units
208,650

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • UDI-DI 00810096150639
  • Lot Numbers: all lots manufactured through 09/01/2026

Distribution

Distributed nationwide across the United States.

Same hazard · inaccurate delivery