Vesco Q Enteral Feeding Pump Recalled for Delivery Inaccuracy
Vesco Medical LLC is recalling 3,493 Vesco Q Enteral Feeding Pumps because they may deliver fluids outside accuracy specifications, potentially causing over- or under-delivery of nutrition.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as Class II, and the source describes a hazard where the pump may deliver fluids outside accuracy specifications, potentially causing temporary or medically reversible adverse health consequences with a remote probability of serious adverse health consequences. The source does not report any illnesses, injuries, or deaths.
Plain-English summary
Vesco Medical LLC is recalling 3,493 units of the Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 through 7.8. The recall covers all serial numbers manufactured through November 5, 2025, and software revisions through 7.8. The affected devices were distributed nationwide in the United States.
The recall was initiated because the feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose if proper controls are not in place. The source states that over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.
The source does not report any illnesses, injuries, or deaths associated with this recall. Consumers and healthcare providers should contact Vesco Medical LLC for instructions on how to proceed with the recall.
The recalled product
- Product
- Vesco Q Enteral Feeding Pump, Model Number VED-701, Software Versions 6.0 - 7.8
- Manufacturer
- Vesco Medical Llc
- Affected units
- 3,493
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- UDI-DI: 00850002550498
- Serial Numbers: all serial numbers manufactured though 11/05/2025
- Software revisions through 7.8
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Vesco Medical Llc
See all →- SevereVesco Q 40mm Screw Cap Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q 500mL Bag Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q 100mL Small Bore Bag Feed Set Recalled for Accuracy Issues
FDA (Devices) · 2026-11-18
- SevereVesco Q 1000mL Bag Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q 1000mL Bag Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
Same hazard · delivery inaccuracy
See all →- HighAlung Technologies Recalls Hemolung Respiratory Assist System Infusion Pumps
FDA (Devices) · 2021-02-10
- SevereCME BodyGuard Microset Infusion Sets Recalled for Delivery Inaccuracy
FDA (Devices) · 2020-07-22
- SevereCME America BodyGuard Microset Infusion Sets Recalled for Delivery Inaccuracy
FDA (Devices) · 2020-07-22
- SevereCME BodyGuard Microset Infusion Sets Recalled for Delivery Inaccuracy
FDA (Devices) · 2020-07-22
- SevereCME America Recalls CMExpress Microbore Sets Due to Delivery Inaccuracy
FDA (Devices) · 2020-07-22
Similar recalls
- SevereVesco Q 1000mL Bag Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q 500mL Bag Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q ENPlus Spike Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q 40mm Screw Cap Feed Set Recalled for Inaccurate Fluid Delivery
FDA (Devices) · 2026-11-18
- SevereVesco Q 100mL Small Bore Bag Feed Set Recalled for Accuracy Issues
FDA (Devices) · 2026-11-18
- HighMOOG Recalls EnteraLite Infinity Enteral Feeding Pumps Due to Voltage Defect
FDA (Devices) · 2012-11-21