Encore Medical Recalls EMPOWR Porous Knee System Tibial Punch
Encore Medical is recalling two units of the EMPOWR Tibial Punch because the locking guide failed to properly position the implant.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device defect that resulted in incorrect implant placement (2-3mm posterior deviation), which poses a risk of harm to patients, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Encore Medical, Lp is recalling two units of the EMPOWR Porous Knee System Tibial Punch, Small, Ref 801-05-202. The product is a non-sterile medical device used in orthopedic surgery.
The recall was initiated because the Locking Punch Guide was not sufficiently guiding the large press fit punch. This defect caused the implant to be placed 2-3mm more posterior than intended.
The affected units, identified by Lot 252564, were distributed to one medical device distributor in California. Healthcare providers and patients should contact Encore Medical for further instructions regarding the return or replacement of these devices.
The recalled product
- Product
- EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.
- Manufacturer
- Encore Medical, Lp
- Affected units
- 2
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Lot 252564
Distribution
Distributed in 1 state:
Part of a larger recall action
This product is one of 4 recalled by Encore Medical, Lp under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Encore Medical, Lp
See all →- ModerateEMPOWR 3D KNEE insertion instruments recalled for incorrect product size packaging
FDA (Devices) · 2026-07-15
- ModerateEMPOWR 3D Knee surgical device recalled due to packaging discrepancy
FDA (Devices) · 2026-07-15
- SevereKnee and Humeral Socket Implants Recalled for Incorrect Labeling
FDA (Devices) · 2026-03-04
- HighShoulder prosthesis humeral socket inserts recalled over incorrect implant labeling
FDA (Devices) · 2026-03-04
- SevereReverse Shoulder Prosthesis Implants Recalled Over Incorrect Labeling
FDA (Devices) · 2026-03-04
Same hazard · improper implant placement
See all →- HighHowmedica Recalls 130 AccuLIF TL Insertion Handles Due to Pivot Defect
FDA (Devices) · 2018-03-07
- HighEncore Medical Recalls DJO EMPOWR Knee Locking Tibial Punch Guide
FDA (Devices) · 2018-01-17
- HighEncore Medical Recalls EMPOWR Porous Knee System Tibial Punch
FDA (Devices) · 2018-01-17
- HighEncore Medical Recalls EMPOWR Porous Knee System Tibial Punch
FDA (Devices) · 2018-01-17
Similar recalls
- HighDJO EMPOWR Tibial Knee Punch Handle Recall Due to Retaining Pin Risk
FDA (Devices) · 2020-10-14
- ModerateEMPOWR 3D Knee Tibial Insert Recalled for Wrong Product in Package
FDA (Devices) · 2025-01-01
- HighKnee tibial inserts recalled over a possible packaging discrepancy
FDA (Devices) · 2026-02-11
- HighKnee Tibial Insert Recalled Due to Component Mix-Up in Packaging
FDA (Devices) · 2023-10-18
- ModerateEncore Medical Recalls EMPOWR PS Knee Inserts Due to Packaging Error
FDA (Devices) · 2017-11-01
- HighEMPOWR VVC Knee Tibial Insert Recalled for Reversed Box Labeling
FDA (Devices) · 2023-03-08