Cardiac stress testing system may assign duplicate exam identifiers to patients
Hillrom Welch Allyn XScribe cardiac stress testing systems may assign duplicate exam identifiers, potentially causing patient data mismatches in medical record systems.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This Class II FDA recall involves a theoretical data-integrity hazard with no reported adverse events or injuries. The risk is that duplicate identifiers could cause patient data mismatches if receiving systems rely solely on UIDs, but actual harm has not been documented, placing this at High severity (3) rather than Severe (4).
Plain-English summary
The Hillrom Welch Allyn XScribe Cardiac Stress Testing System is a programmable diagnostic computer used to conduct cardiac stress tests. Approximately 766 units of this system have been distributed worldwide.
The systems may assign duplicate Unique Identifiers (UIDs) to exam files. This could pose a problem if the receiving medical system (such as a Picture Archiving and Communication System, or PACS) relies solely on the UID to match exams to patient records, potentially resulting in mismatches between patient identification and their physiological test data.
This recall affects healthcare facilities and cardiac testing centers using the affected XScribe systems with software versions V6.0.0 or V6.3.2 released July 2, 2015.
The recalled product
- Product
- Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
- Manufacturer
- Baxter Healthcare Corporation
- Affected units
- 766
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Software Versions: Software Version V6.0.0 V6.3.2
- Release Date 02-Jul-2015 Units: 1) UDI/DI 00732094259032
- Serial Numbers: 115340269178
- 117060553883
- 117060553884
- 117060553885
- 119230003169
- 2) UDI/DI 00732094259070
- Serial Numbers: 117130559202
- 117130559203
- 117130559204
- 3) UDI/DI 00732094259100
- Serial Numbers: 116270497046
- 117190567440
- 4) UDI/DI 00732094259148
- Serial Numbers: 117350583961
- 119060000843
- 122090000529
- 5) UDI/DI 00732094259179
- Serial Numbers: 116240458434
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 6 recalled by Baxter Healthcare Corporation under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · Baxter Healthcare Corporation
See all →- HighBaxter Recalls Likorall Overhead Lifts Due to Motor Shaft Breakage Risk
FDA (Devices) · 2026-10-21
- ModerateBaxter Recalls Two VOLARA Systems Due to Potential Flow Rate Inaccuracy
FDA (Devices) · 2026-09-30
- ModerateBaxter PRO+ 36-inch Mattress Recalled for UDI Labeling Error
FDA (Devices) · 2026-09-30
- ModerateBaxter PRO+ Mattress Recalled for UDI Labeling Error
FDA (Devices) · 2026-09-30
- ModerateBaxter Recalls PRO+ MRS Mattress Due to UDI Labeling Error
FDA (Devices) · 2026-09-30
Same hazard · duplicate identifier
See all →- HighHillrom Diagnostic Cardiology Suite Software Duplicate UID Assignment Recall
FDA (Devices) · 2024-11-13
- HighHillrom HScribe Holter System cardiac monitoring devices duplicate identifier risk
FDA (Devices) · 2024-11-13
- HighElectrocardiograph DICOM Files Risk Patient Data Mismatch Due to Duplicate Identifiers
FDA (Devices) · 2024-11-13
Similar recalls
- HighXScribe Cardiac Stress Testing System may distort ECG readings
FDA (Devices) · 2024-01-17
- HighCardiac Stress Testing Device May Produce Distorted ECG Readings
FDA (Devices) · 2024-01-17
- HighCardiac assessment system may display test reports in wrong patient records
FDA (Devices) · 2021-06-30
- HighWelch Allyn Q-Stress and XScribe Diagnostic Devices Recalled for Safety Standard Deficiency
FDA (Devices) · 2020-05-06
- HighCardiac Stress Testing System may produce distorted electrocardiogram readings
FDA (Devices) · 2024-01-17
- HighSensis Vibe Cardiac Imaging System Software May Crash During Documentation
FDA (Devices) · 2024-07-31