The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0319-2027·Announced 2026-11-11

Natus Medical Recalls Embletta MPR PG Sleep Monitors Due to Memory Formatting Issues

Natus Medical is recalling 288 Embletta MPR PG Sleep Monitors because internal memory formatting errors may cause incorrect patient data or incomplete studies.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The hazard is a software defect causing incorrect patient data or incomplete diagnostic recordings. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Natus Medical Incorporated is recalling 288 units of the Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number 2003011. The devices were distributed worldwide, including in the United States, Canada, China, Italy, and Portugal.

The recall is due to a defect where the device may be unable to correctly format its internal memory. This issue can cause previously recorded studies to remain on the device and be updated with incorrect patient information from a subsequent study. Additionally, the defect may prevent studies from initiating recording successfully or cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.

The source text does not report any illnesses, injuries, or deaths associated with this recall.

The recalled product

Product
Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011
Affected units
288

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI-DI: 00382830023764

Distribution

Distributed nationwide across the United States.