Infusion line packs recalled over incomplete seals in sterile pouches
Alcon Research is recalling 25,778 accessory devices for Unity and Constellation surgical consoles because of incomplete seals in the pouch that provides the sterile barrier.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: a failed sterile barrier on a surgical accessory creates an infection risk, and no injuries are reported.
Plain-English summary
Accessory devices for Alcon Unity and Constellation surgical consoles are being recalled. The affected items identified in the record include catalog number 8065750918, an irrigation and aspiration line pack, and 8065750957, a Constellation ship assembly, with the firm referring to a fuller affected product list.
The recall follows incomplete seals in the pouch which provides the sterile barrier.
The recall covers 25,778 units, including UDI-DI 00380657509188 from lot 1790L4. Distribution was worldwide, including nationwide in the United States. The FDA has classified the action as Class II.
An incomplete seal means the sterile barrier may not have held. Facilities holding affected lots should quarantine them and follow Alcon's instructions.
The recalled product
- Product
- Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065750918 PAK,I/A LINE,NGVS 8065750957 ASSY,SHIP,CONSTELLATION VFC 8065750972…
- Manufacturer
- Alcon Research LLC
- Affected units
- 25,778
Distribution
Part of a larger recall action
This product is one of 5 recalled by Alcon Research LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · Alcon Research LLC
See all →- SevereGenTeal Tears Lubricant Eye Gel Recalled Due to Sterility Concerns
FDA (Drugs) · 2026-05-06
- SevereSystane Lubricant Eye Gel Recalled Due to Sterility Concerns
FDA (Drugs) · 2026-05-06
- CriticalOphthalmic surgical packs recalled over incomplete seals affecting sterility
FDA (Devices) · 2025-12-31
- ModerateVitrectomy probe packs recalled with no reason stated in the record
FDA (Devices) · 2025-12-03
- HighEntry system accessories recalled over incomplete seals in sterile pouches
FDA (Devices) · 2025-11-05
Same hazard · seal failure
See all →- SevereReprocessed Electrophysiology Catheter with Incomplete Sterile Seals Recalled
FDA (Devices) · 2026-05-20
- HighMedline dialysis maintenance kits recalled over Tego connector seal defects
FDA (Devices) · 2026-03-04
- HighMedline apheresis and central line kits recalled over connector seal defects
FDA (Devices) · 2026-03-04
- HighMedline dialysis kits recalled over damaged Tego connector silicone seals
FDA (Devices) · 2026-03-04
- HighMedline dialysis dressing change kits recalled over connector seal defects
FDA (Devices) · 2026-03-04
Similar recalls
- HighAlcon Constellation procedure packs may have damaged lidding compromising sterility
FDA (Devices) · 2025-07-16
- CriticalOphthalmic surgical packs recalled over incomplete seals affecting sterility
FDA (Devices) · 2025-12-31
- HighAlcon Constellation Vision System 20 GA TOTALPLUS Combined Procedure Pak Recall
FDA (Devices) · 2020-04-29
- HighAlcon Constellation vitrectomy packs may have damaged lidding compromising sterility
FDA (Devices) · 2025-07-16
- HighAlcon Recalls Eye-Pak 7407 Tray Support Covers Due to Seal Issues
FDA (Devices) · 2014-10-01
- ModerateAlcon Accurus Vitreous Probes Recalled for Packaging Seal Defect
FDA (Devices) · 2015-07-08