The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0241-2019·Announced 2018-10-31

Custom Healthcare Systems Recalls Injection Trays Due to Sterility Compromise

Custom Healthcare Systems is recalling 40 units of CHS Custom Convenience Kit-INJECTION TRAY because incorrect sealing specifications compromised product sterility.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II and involves a risk-of-harm product (compromised sterility) where no specific injuries or illnesses have been reported in the source text.

Plain-English summary

Custom Healthcare Systems, Inc. is recalling 40 units of the CHS Custom Convenience Kit-INJECTION TRAY (Product Number: B9-17201). The recall involves Lot Number 70518 with an expiration date of 2/1/2021. The product was distributed to facilities in Connecticut, Georgia, Kentucky, North Carolina, Virginia, Maryland, Oklahoma, Illinois, and South Carolina.

The recall was initiated because the convenience kits were sealed in immediate Tyvek pouches using incorrect specifications. This error compromised the sterility of the product. The U.S. Food and Drug Administration has classified this as a Class II recall.

Consumers and healthcare facilities should stop using the affected injection trays immediately. Users should contact Custom Healthcare Systems, Inc. for instructions on how to return the product or obtain a replacement.

The recalled product

Product
CHS Custom Convenience Kit-INJECTION TRAY Product Number: B9-17201
Affected units
40

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • Lot Number: 70518
  • Exp. Date: 2/1/2021

Distribution

Distributed in 9 states:

Part of a larger recall action

This product is one of 29 recalled by Custom Healthcare Systems, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 29 products in this recall →