Agency classClass IIOur severity:HighFDA (Devices)·Z-0239-2024·Announced 2023-11-15

Disposable Spetzler-Malis Bipolar Forceps Recalled for Incorrect Shelf Life Label

Stryker is recalling 2,321 disposable Spetzler-Malis Bipolar forceps due to a labeling error showing an incorrect shelf life. The product label indicates 54 months, but the actual shelf life is 36 months.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: This is an FDA Class II recall of a surgical instrument due to a shelf-life labeling error. No illnesses or injuries have been reported, but using surgical instruments beyond their intended shelf life constitutes a risk-of-harm scenario that meets the rubric criterion for High severity.

Plain-English summary

Stryker Corporation is recalling 2,321 units of disposable Spetzler-Malis Bipolar forceps (18CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-180-005). The affected lot is 200507002 (GTIN: 7613327300192).

An error was identified on the product label. The label indicates a shelf life of 54 months (4.5 years), but the actual shelf life is 36 months (3 years). This discrepancy means products labeled with these incorrect expiration dates could potentially be used beyond their actual shelf life.

The affected products were distributed nationwide in the United States and internationally to Australia, Canada, Japan, Korea, and Singapore. Healthcare facilities and users who have received these products should verify the correct 36-month shelf life and contact Stryker for further guidance.

The recalled product

Product
Disposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-180-005
Affected units
2,321

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (2)

  • GTIN: 7613327300192
  • Lot Numbers: 200507002

Distribution

Distributed nationwide across the United States.