HighFDA (Devices)·Z-0231-2026·Announced 2025-10-29

[pending] MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electroc

Pending LLM rewrite. Source: FDA_DEVICE Z-0231-2026.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Pending LLM scoring against the rubric.

Plain-English summary

If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.

The recalled product

Product
MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4
Manufacturer
GE Medical Systems, LLC

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI-DI: 00840682125499

Distribution

Distributed nationwide across the United States.