ECG trunk cables recalled because improper module placement can cause overheating
GE is recalling 29,225 MAC VU360 acquisition trunk cables and module holders because incorrect placement of the module can wear the cable until it overheats in use.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: a worn cable that can overheat in patient use is a burn and electrical hazard, and no injuries are reported.
Plain-English summary
The MAC VU360 Acquisition Trunk Cable and Module Holder, used with a high performance multichannel resting electrocardiograph, model number 2030360-001, product codes VU2 and VU4, is being recalled.
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable, visible as a damaged sheath or exposed wires. If a damaged cable is not recognized and removed from service, it can overheat with continued use.
The recall covers 29,225 units carrying UDI-DI 00840682125499. Distribution was worldwide, including nationwide in the United States and to Australia, Austria and Bahrain among other countries. The FDA has classified the action as Class II.
The enforcement record describes the overheating pathway and reports no injuries. Sites should inspect cables for a damaged sheath or exposed wires and follow GE's instructions.
The recalled product
- Product
- MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4
- Manufacturer
- GE Medical Systems, LLC
- Category
- Medical Device — ECG accessory
- Affected units
- 29,225
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI: 00840682125499
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · GE Medical Systems, LLC
See all →- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- HighGE Medical Optima XR200/220/240amx Mobile Radiographic Systems Recall
FDA (Devices) · 2026-07-22
- HighGiraffe OmniBed incubator canopy fasteners can fall into patient compartment
FDA (Devices) · 2026-06-24
- HighGiraffe OmniBed replaceable units share the canopy fastener disengagement defect
FDA (Devices) · 2026-06-24
Same hazard · cable damage
See all →- High
- High
- HighMagstim Recalls Neurosign V4 Intraoperative Nerve Monitors Due to Cable Damage Risk
FDA (Devices) · 2020-09-02
- HighSiemens Artis zee Ceiling Systems Recalled for Cable Damage Risk
FDA (Devices) · 2014-03-05
Similar recalls
- ModerateGE Healthcare Recalls MAC VU360 Electrocardiographs for Patient Identification Errors
FDA (Devices) · 2020-12-16
- HighGE Healthcare Recalls MAC VU360 Electrocardiographs Due to Patient ID Errors
FDA (Devices) · 2019-07-31
- ModerateGE Healthcare Recalls ECG Trunk Cables Due to Defibrillation Risk
FDA (Devices) · 2019-07-03
- HighGE Healthcare Recalls Integrated ECG Cables Due to Short Circuit Risk
FDA (Devices) · 2019-04-03
- SevereGE Healthcare MAC VU360 Electrocardiograph Software Recall
FDA (Devices) · 2026-08-12
- HighPatient Warming Mattress Recalled for Cable Connection Defect
FDA (Devices) · 2025-02-26