Blood Lead Test Kits Recalled for Defective Treatment Reagent Caps
Magellan Diagnostics recalls LeadCare II Blood Lead Test Kits nationwide due to defective plastic caps on reagent tubes. A small percentage of caps may not seal properly, potentially leading to inaccurate test results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II medical device recall with no reported illnesses or injuries. The defect creates a risk of inaccurate diagnostic results through inadequate seals on reagent tube caps, meeting the rubric criterion for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Magellan Diagnostics, Inc. is recalling 11,442 LeadCare II Blood Lead Test Kits (549,216 tests) distributed nationwide. The test kits are used for in vitro diagnostic measurement of lead in capillary whole blood.
During an internal audit, the manufacturer identified that a small percentage (less than 10 percent) of the plastic caps used to seal treatment reagent tubes do not meet minimum dimensional requirements. These defective caps may form inadequate seals.
Inadequate cap seals may allow leaking of hydrochloric acid or evaporation of the treatment reagent. Either condition changes the volume and concentration of the reagent, altering the ratio of patient sample or control material to reagent. This can result in falsely elevated test results, including false positive readings or out-of-range control results.
Healthcare providers and laboratories using the affected lot numbers should discontinue use and contact Magellan Diagnostics for replacement or further instructions.
The recalled product
- Product
- LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
- Manufacturer
- Magellan Diagnostics, Inc.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Magellan Diagnostics, Inc.
See all →- ModerateLeadCare II Blood Lead Test Kit Recall for Inaccurate Results
FDA (Devices) · 2026-04-29
- HighLeadCare Plus Blood Lead Test Kit Recalled for Unexpectedly Elevated Results
FDA (Devices) · 2026-04-29
- ModerateLeadCare Ultra Blood Lead Test Kit Recalled for False Positive Results
FDA (Devices) · 2026-04-29
- HighBlood lead test kits recalled over mislabeled calibration buttons
FDA (Devices) · 2025-12-03
- SevereLeadCare Ultra Blood Lead Test Kit Recalled for Control Test Inaccuracy
FDA (Devices) · 2021-07-07
Same hazard · seal failure
See all →- SevereReprocessed Electrophysiology Catheter with Incomplete Sterile Seals Recalled
FDA (Devices) · 2026-05-20
- HighMedline dialysis dressing change kits recalled over connector seal defects
FDA (Devices) · 2026-03-04
- HighMedline dialysis maintenance kits recalled over Tego connector seal defects
FDA (Devices) · 2026-03-04
- HighMedline apheresis and central line kits recalled over connector seal defects
FDA (Devices) · 2026-03-04
- HighMedline dialysis kits recalled over damaged Tego connector silicone seals
FDA (Devices) · 2026-03-04
Similar recalls
- SevereMagellan LeadCare Blood Lead Test Kits Recalled for Underestimating Results
FDA (Devices) · 2017-06-14
- SevereLeadCare II Blood Lead Test Kit Control Test Inaccuracy Recall
FDA (Devices) · 2021-07-07
- HighMagellan LeadCare II Blood Lead Test Kit Recalled for Calibration Defect
FDA (Devices) · 2018-11-07
- ModerateLeadCare II Blood Lead Test Kit Recall for Inaccurate Results
FDA (Devices) · 2026-04-29
- HighBlood lead test kits recalled over mislabeled calibration buttons
FDA (Devices) · 2025-12-03
- HighMagellan LeadCare II Blood Lead Test Kit Recall for Out-of-Range Controls
FDA (Devices) · 2017-08-23