DePuy Mitek Screw and Washer Depth Gauge Recalled for Design Offset
DePuy Mitek is recalling 401 Mitek Screw and Washer Depth Gauges because a design offset causes readings to be approximately 3.2mm shorter than the actual bone hole length.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a medical device with a design defect that affects measurement accuracy, representing a risk of harm where injury has not yet been reported.
Plain-English summary
DePuy Mitek, Inc., a Johnson & Johnson Co., is recalling 401 units of the Mitek Screw and Washer Depth Gauge, Product Code 219024. The recall is due to a design offset in the device that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
The affected products were distributed nationwide in the United States to the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, and WI. The products were also distributed to foreign countries including Australia, Canada, Chile, China, Chechnya, Denmark, Italy, Japan, Jordan, Netherlands, Poland, Slovenia, Sweden, Turkey, and the UK.
The agency has classified this recall as Class II. No specific instructions for consumers or healthcare providers are provided in the source text regarding immediate actions to take.
The recalled product
- Product
- Mitek Screw and Washer Depth Gauge, Product Code 219024
- Manufacturer
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Affected units
- 401
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI: 10886705005390
Distribution
Related recalls
Same manufacturer · DePuy Mitek, Inc., a Johnson & Johnson Co.
See all →- HighCOR Disposable Cartilage Transplant Kit recalled for missing graft loader pin
FDA (Devices) · 2025-03-19
- HighCartilage Transplant System Disposable Kits Recalled for Missing Component
FDA (Devices) · 2025-03-19
- HighSurgical Anchor Inserter Tip Fragmentation During Patient Procedures
FDA (Devices) · 2023-06-14
- HighDePuy Mitek surgical anchor inserter may fracture during use
FDA (Devices) · 2023-06-14
- HighSurgical anchor inserter recalled for potential tip fracture and bone damage
FDA (Devices) · 2023-06-14
Same hazard · design defect
See all →- HighUnited Orthopedic Recalls U2 Total Knee System Femoral Offset Bushings
FDA (Devices) · 2026-08-12
- SevereBioFlo DuraMax catheters recalled as Class I over sheath introducer defect
FDA (Devices) · 2026-04-01
- SevereMerit 16F splittable sheath introducer recalled as Class I bulk product
FDA (Devices) · 2026-04-01
- SevereProGuide chronic dialysis catheters recalled as Class I over sheath introducer defect
FDA (Devices) · 2026-04-01
- SevereCentrosFLO hemodialysis catheters recalled as Class I over sheath introducer defect
FDA (Devices) · 2026-04-01
Similar recalls
- ModerateDePuy Mitek Recalls FMS 3.5mm Ultra-Aggressive Cutter Shaver Blades
FDA (Devices) · 2016-01-27
- ModerateDePuy Mitek Recalls FMS neXtra 3.5mm Cutter Shaver Blades Due to Labeling Error
FDA (Devices) · 2016-01-27
- ModerateDePuy Mitek Recalls FMS neXtra 3.5mm Aggressive Cutter Shaver Blades
FDA (Devices) · 2016-01-27
- ModerateDePuy Mitek Sterile Reamers Recalled for Incorrect Size Labeling
FDA (Devices) · 2012-12-26
- HighDePuy Mitek LATARJET Extraction Driver Recalled for Breakage Risk
FDA (Devices) · 2017-06-28
- ModerateMedtronic Recalls Vertex Select Multi Axial Screws Due to Length Discrepancy
FDA (Devices) · 2020-05-20