PICC and gastrostomy tube packs recalled after being sterilized more than once
American Contract Systems is recalling PICC G-Tube convenience packs that were re-gassed after a sterilization nonconformance, without validation for multiple cycles.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: quality and safety cannot be confirmed for kits used in invasive procedures, and no injuries are reported.
Plain-English summary
Medical convenience kits sold as PICC G-TUBE PACK, item number UDPG88W, are being recalled.
The identified products were re-gassed, meaning sterilized more than once, after a nonconformance occurred during the initial ethylene oxide gas injection process. These products have not been validated for exposure to multiple sterilization cycles, so product quality, safety and effectiveness cannot be confirmed.
The action covers 32,433 units in total across the firm's affected kits. This item carries UDI-DI 191072216359 and lot 7935611. The product was distributed in South Dakota, Iowa, Minnesota, Washington and Illinois. The FDA has classified the action as Class II.
Facilities should identify affected item, lot and bag serial numbers, quarantine them and follow the firm's instructions.
The recalled product
- Product
- Medical convenience kits UDPG88W PICC G-TUBE PACK-230209
- Manufacturer
- American Contract Systems Inc.
- Category
- Medical Device — Surgical kit
- Affected units
- 32,433
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- then whole lot is affected
Distribution
Part of a larger recall action
This product is one of 18 recalled by American Contract Systems Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 18 products in this recall →Related recalls
Same manufacturer · American Contract Systems Inc.
See all →- SevereFDA Recalls Medical Kits with Contaminated Pericare Wipes
FDA (Devices) · 2026-10-07
- HighFDA Recalls Medical Kits with Contaminated Pericare Wipes
FDA (Devices) · 2026-10-07
- HighRecall of Medical Convenience Kits with Contaminated Pericare Wipes
FDA (Devices) · 2026-10-07
- ModerateAmerican Contract System Convenience Kits Recalled for Performance Failures
FDA (Devices) · 2026-09-02
- ModerateAmerican Contract Systems Recalls Convenience Kits with Non-Conforming Catheters
FDA (Devices) · 2026-09-02
Same hazard · resterilization
See all →- HighGeneral and laparotomy packs recalled after being sterilized more than once
FDA (Devices) · 2025-10-22
- HighCardiovascular procedure packs recalled after being sterilized more than once
FDA (Devices) · 2025-10-22
- HighDilation and curettage packs recalled after being sterilized more than once
FDA (Devices) · 2025-10-22
- HighCardiac catheterization packs recalled after being sterilized more than once
FDA (Devices) · 2025-10-22
- HighCentral line insertion kits recalled after being sterilized more than once
FDA (Devices) · 2025-10-22
Similar recalls
- HighSurgical instrument convenience packs recalled for inadequate sterilization validation
FDA (Devices) · 2022-09-14
- HighPICC and Central Line Packs Exposed to Unvalidated Sterilization Cycles
FDA (Devices) · 2022-09-14
- HighMedical device convenience packs exposed to unvalidated sterilization cycles
FDA (Devices) · 2022-09-14
- HighObstetric and gynecology surgical packs recalled for unvalidated sterilization exposure
FDA (Devices) · 2022-09-14
- HighCysto Packs Recalled Due to Inadequate Sterilization Validation
FDA (Devices) · 2022-09-14
- HighMedical Procedure Kit Recalled Over Potential Sterilization Compromise
FDA (Devices) · 2024-08-14