Covidien Recalls Puritan Bennett 980 Ventilators Due to Capacitor Failure
Nellcor Puritan Bennett is recalling 324 Puritan Bennett 980 Ventilators because a capacitor component can fail, causing a dim screen and thermal odor.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor."
Plain-English summary
Nellcor Puritan Bennett Inc. (dba Covidien LP) is recalling 324 units of the Puritan Bennett 980 Ventilator System. The recall is due to a defect in the blacklight inverter PCBA where a capacitor component can fail.
This failure can cause the device screen to become dim, though it remains functional, and may produce a thermal odor. The affected units were distributed worldwide, including in the United States, Canada, Mexico, Saudi Arabia, South Africa, and the United Arab Emirates.
Healthcare facilities and users should check their inventory for the specific model numbers listed in the recall notice. If a unit is affected, it should be removed from service and returned to the manufacturer for repair or replacement.
The recalled product
- Product
- Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU.
- Manufacturer
- Nellcor Puritan Bennett Inc. (dba Covidien LP)
- Category
- Medical Device — Ventilator
- Affected units
- 324
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model number PB980 Ventilator:35B1400143
- 35B1400684
- 35B1400627
- 35B1400708
- 35B1400398
- 35B1400482
- 35B1400491
- 35B1400512
- 35B1400601
- 35B1400121
- 35B1400568
- 35B1400590
- 35B1400595
- 35B1400599
- 35B1400606
- 35B1400106
- 35B1400146
- 35B1400169
- 35B1400226
- 35B1400260
Distribution
Related recalls
Same manufacturer · Nellcor Puritan Bennett Inc. (dba Covidien LP)
See all →- SevereFDA Recalls 324 Puritan Bennett 980 Ventilators Due to Software Issue
FDA (Devices) · 2014-11-05
- HighVital Sync Virtual Patient Monitoring Platform software recall for alarm mismatch
FDA (Devices) · 2014-08-20
- ModerateVital Sync Virtual Patient Monitoring Platform Recalled for Alarm Priority Mismatch
FDA (Devices) · 2014-08-20
- HighPuritan Bennett 840 Ventilator Recall for Display Loss
FDA (Devices) · 2014-07-02
- HighNellcor Puritan Bennett PB840 Ventilator Display Component Recalled
FDA (Devices) · 2014-06-25
Same hazard · component failure
See all →- SevereAbiomed Recalls Impella Optical Controllers Due to Purge Flag Component Failure
FDA (Devices) · 2026-09-23
- ModerateSpacelabs Xhibit Central Station Recalled for DisplayPort Audio/Video Failures
FDA (Devices) · 2026-08-19
- High
- SevereKawasaki Teryx4 and Teryx5 H2 drive converter sheaves can break
CPSC · 2026-06-11
- HighOlympus Thunderbeat II Shears Ultrasonic Mode distal tip detachment risk
FDA (Devices) · 2026-04-29
Similar recalls
- SeverePuritan Bennett 980 Series Ventilator Recalled for Capacitor Assembly Defect
FDA (Devices) · 2021-12-15
- SeverePuritan Bennett 980 Series Ventilator Recall Due to Assembly Defect
FDA (Devices) · 2021-12-15
- SeverePuritan Bennett 980 Series Ventilator Recalled for Manufacturing Assembly Defect
FDA (Devices) · 2021-12-15
- SeverePuritan Bennett 980 Ventilator Recalled for Capacitor Assembly Error
FDA (Devices) · 2021-12-15
- SeverePuritan Bennett 980 Ventilators Recalled Due to Manufacturing Assembly Error
FDA (Devices) · 2021-12-15
- SeverePuritan Bennett 980 Series Ventilator Recalled for Assembly Defect
FDA (Devices) · 2021-12-15