Abbott Recalls Emerald Diluent Tubing Due to Cloudiness and Film
Abbott Laboratories is recalling 22 CELL-DYN Emerald analyzers with diluent tubing that appears cloudy or has a film, which may cause falsely elevated platelet results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a risk of harm (falsely elevated platelet results) where no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Abbott Laboratories is recalling 22 CELL-DYN Emerald analyzers equipped with Emerald Diluent Tubing (List number 09H50-01). The recall involves specific instrument serial numbers and tubing lot #2018-10-16. The affected products were distributed to laboratories in Georgia, Louisiana, Massachusetts, North Carolina, New York, South Carolina, and Texas, as well as to locations in Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.
The recall was initiated because the diluent tubing appears cloudy or has a powder/film on the inner and outer surfaces. This condition may result in falsely elevated platelet results. The agency has classified this as a Class II recall.
Users of the affected instruments should contact Abbott Laboratories for instructions on how to proceed. The recall number is Z-0173-2020.
The recalled product
- Product
- Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.
- Manufacturer
- Abbott Laboratories
- Hazard
- Affected units
- 22
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Tubing lot #2018-10-16
- Instrument serial numbers 007689
- 007690
- 007691
- 007692
- 007705
- 007706
- 007707
- 007708
- 007709
- 007728
- 007694
- 007695
- 007697
- 007766
- 007767
- 007768
- 007765
- 007710
- 007699
Distribution
Related recalls
Same manufacturer · Abbott Laboratories
See all →- HighChemistry analyzer control modules recalled over software that may give erroneous results
FDA (Devices) · 2025-12-24
- HighHematology analyzers recalled because saturation flags may not be applied
FDA (Devices) · 2025-11-19
- HighCELL-DYN Ruby hematology analyzer system expiration date validation
FDA (Devices) · 2025-04-30
- HighCELL-DYN Ruby Hematology Analyzer Expiration Date Software Defect
FDA (Devices) · 2025-04-30
- HighCELL-DYN Ruby hematology analyzer expiration date software defect
FDA (Devices) · 2025-04-30
Same hazard · inaccurate results
See all →- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
Similar recalls
- ModerateAbbott Recalls CELL-DYN Emerald Cleaner Due to Analyzer Quality Control Issues
FDA (Devices) · 2016-04-27
- ModerateAbbott Recalls CELL-DYN Emerald Analyzers Due to Quality Control Errors
FDA (Devices) · 2019-04-10
- HighAbbott Hematology Analyzers Recalled for Unlabeled Latex Component
FDA (Devices) · 2023-12-27
- ModerateAbbott Recalls CELL-DYN Ruby Hematology Analyzers Due to Pump Relay Failure
FDA (Devices) · 2017-03-15
- HighAbbott Activated Alanine Aminotransferase Reagent Recalled for Inaccurate Test Results
FDA (Devices) · 2021-08-11
- HighAlinity c Hemoglobin Reagent Kit Recall Due to Potentially Falsely Elevated Results
FDA (Devices) · 2022-08-03